- 2026-04-08FDA-DeviceCivco Medical Instruments Co. Inc.Class IIStop using Civco eTRAX Needle System Starter Kit
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using Medline Anterior Hip Pack DYNJ64672B
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIMedline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTO
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using affected Medline bronchoscopy kits
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using recalled Medline respiratory kits
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using affected Medline thoracic robot kits
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using affected Medline bronchoscopy kits immediately
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using Medline Airway Exam Kit DYKE1796
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- 2026-04-08FDA-DeviceSUMMA THERAPEUTICS, LLCClass IIStop using Finesse BTK Multicath catheters
Potential for the balloon in the device to not meet burst specifications.
- 2026-04-08FDA-DeviceSynthes (USA) Products LLCClass IIStop using Synthes VOLT Cortex Screw lot 68982P8
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
- 2026-04-08FDA-DeviceSynthes (USA) Products LLCClass IICheck Synthes bone tap lot 68900P9 immediately
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IIStop using affected da Vinci cautery hook instruments
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IICheck da Vinci Tenaculum Forceps for cable damage
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IIStop using affected da Vinci grasping retractors
Due to increased complaints for broken/frayed grip cables for reusable instruments.
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IIStop using da Vinci Mega Needle Driver instruments
Due to increased complaints for broken/frayed grip cables for reusable instruments.
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IIStop using affected da Vinci grasper instruments
Due to increased complaints for broken/frayed grip cables for reusable instruments.
- 2026-04-08FDA-DeviceIntuitive Surgical, Inc.Class IIStop using recalled da Vinci scissors instruments
Due to increased complaints for broken/frayed grip cables for reusable instruments.
- 2026-04-08FDA-DeviceOlympus Corporation of the AmericasClass IIStop using affected Olympus SOLTIVE Pro laser units
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
- 2026-04-08FDA-DeviceOlympus Corporation of the AmericasClass IIStop using affected Olympus SOLTIVE laser systems
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using Medline radiology kits lot 25GBB924
There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.
- 2026-04-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IICheck your Philips Vue Motion V12 software
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using Medline ENFIT G-tube connector kits
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
- 2026-04-08FDA-DeviceMedline Industries, LPClass IIStop using Medline ENFIT G-tube connectors
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
- 2026-04-08FDA-DeviceUltradent Products, Inc.Class IIStop using Ultradent Jiffy polisher cups
Composite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment.
- 2026-04-08FDA-DeviceStryker CommunicationsClass IICheck Stryker CHROMOPHARE surgical light ceiling support
Surgical light assembly may not adequate support the weight of the ceiling cover.
- 2026-04-08FDA-DeviceMentor Texas, LP.Class IIStop using Mentor Artoura tissue expanders
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
- 2026-04-08FDA-DeviceMentor Texas, LP.Class IIStop using Mentor Artoura Breast Tissue Expanders
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
- 2026-04-08FDA-DeviceMentor Texas, LP.Class IIStop using Mentor Artoura Breast Tissue Expanders
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
- 2026-04-08FDA-DeviceMentor Texas, LP.Class IIStop using Mentor CPX 4 breast tissue expanders
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
- 2026-04-08FDA-DeviceMentor Texas, LP.Class IIStop using Mentor CPX 4 tissue expanders
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.