- 2024-09-18FDA-DeviceSmiths Medical ASD Inc.Class IPneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX
There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
- 2024-09-18FDA-DeviceSmiths Medical ASD Inc.Class IPneupac paraPAC plus 300 ventilator kit, REF P300NXX
There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
- 2024-09-18FDA-DeviceB Braun Medical IncClass IInfusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.
- 2024-09-18FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-645L
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury
- 2024-09-18FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-640L
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury
- 2024-09-11FDA-DeviceBeckman Coulter Biomedical GmbHClass IIPE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2024-09-11FDA-DeviceBeckman Coulter Biomedical GmbHClass IIPE Centrifuge, REF: B36365, a component of the Power Express
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2024-09-11FDA-DeviceBeckman Coulter Biomedical GmbHClass IIPower Express, REF B90918
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 with SmartAssist S2 Set, EU; Product Code: 1000482;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 SmartAssist S2 Set, JP; Product Code: 1000459;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Smart Assist Set; Product Code: 1000402;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 Set AU; Product Code: 1000361;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 SmartAssist Set, JP; Product Code: 1000211;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 with SmartAssist S2 Set, US; Product Code: 1000100;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIPump 381 Pump Set (US); Product Code: 1000080;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 with SmartAssist Set, US; Product Code: 0550-0008;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Smart Assist Set APAC; Product Code: 0048-0047;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIPump 371 14F LT CMR Set; Product Code: 0048-0045;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Smart Assist Set, Canada; Product Code: 0048-0044;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Pump Set, Japan; Product Code: 0048-0034-JP;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Smart Assist Set, JP; Product Code: 0048-0024-JP;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DeviceAbiomed, Inc.Class IIImpella CP Smart Assist Set, EU; Product Code: 0048-0014;
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- 2024-09-11FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIPhilips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
- 2024-09-11FDA-DeviceMindray DS USA, Inc. dba Mindray North AmericaClass IIDesflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number115-066759-00
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
- 2024-09-11FDA-DeviceMindray DS USA, Inc. dba Mindray North AmericaClass IIIsoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number 115-066758-00
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
- 2024-09-11FDA-DeviceMindray DS USA, Inc. dba Mindray North AmericaClass IISevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention