Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2024-05-29FDA-DeviceOptumHealth Care Solutions LLCClass IOptum Nimbus II Plus, Model number AAA-00170; Ambulatory Infusion Pump
Nimbus II Plus, Infusion Pump Systems, manufactured by InfuTronix, are being removed the from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing.
- 2024-05-29FDA-DeviceJiangsu Shenli Medical Production Co., Ltd.Not Yet ClassifiedCheck Medline syringe lot codes; contact supplier
Piston syringes sizes and configurations are out of the range of devices cleared under the firm's 510(k).
- 2024-05-22FDA-DeviceCardinal Health 200, LLCClass IICardinalHealth Curity Flexible Adhesive Bandage X-Large 2 x 3.75 in. (5.1 x 9.5 cm), Model 44102, Sterile, Hypoallergenic.
Potential contamination of the product with latex adhesive residual.
- 2024-05-22FDA-DeviceMacherey Nagel Gmbh & Co. KgClass IIReagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.
The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.
- 2024-05-22FDA-DeviceCook Medical IncorporatedClass IIEchoTip¿ Ultra Endoscopic Ultrasound Access Needle, ECHO-19, G31520
Devices may contain elevated levels of bacterial endotoxin.
- 2024-05-22FDA-DeviceCook Medical IncorporatedClass IIEchoTip¿ Ultra HD Endoscopic Ultrasound Access Needle, ECHO-HD-19-A, G52012
Devices may contain elevated levels of bacterial endotoxin.
- 2024-05-22FDA-DeviceMaquet Medical Systems USAClass IICARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
- 2024-05-22FDA-DevicePhilips North America LlcClass IIIngenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
- 2024-05-22FDA-DevicePhilips North America LlcClass IIIngenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
- 2024-05-22FDA-DeviceAndover Healthcare Inc.Class IISterile CoFlex NL: Latex Free Flexible Cohesive Bandage, Tan, 4INx5YD (5400S)
A packaging seal gap could impact the sterility of the product.
- 2024-05-22FDA-DevicePhilips North America LlcClass IIPatient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
Event Catalog information does not save when copied and transferred from one unit to another.
- 2024-05-22FDA-DeviceSophysa S.A.Class IIThe Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VTT
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
- 2024-05-22FDA-DeviceSophysa S.A.Class IIThe Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VT
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
- 2024-05-22FDA-DeviceLeMaitre Vascular, Inc.Class IISyntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
The guide tip can become damaged and result in the tip detaching.
- 2024-05-22FDA-DeviceFisher & Paykel Healthcare, Ltd.Class IIAirvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
- 2024-05-22FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINovalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation and CO2 removal) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure Product Code: F30000163
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.
- 2024-05-22FDA-DeviceMedartis Inc.Class IIAPTUS 2.8 TriLock Screw 22mm, HD7, 1/Pkg. Metal bone fixation fastener used with the APTUS fixation system.
Screw is 16mm long instead of 22mm long.
- 2024-05-22FDA-DeviceHeartSine Technologies LtdClass IIHeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.
- 2024-05-22FDA-DeviceQUIDELORTHOClass IIVITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820
Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.
- 2024-05-22FDA-DeviceQUIDELORTHOClass IIVITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 131 5589
T3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.
- 2024-05-22FDA-DeviceBaxter Healthcare CorporationClass IISpectrum IQ Infusion pump, Product Code 357009
There is a potential for cracks on the mount of the front panel of the device.
- 2024-05-22FDA-DeviceKarl Storz EndoscopyClass IILASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000
Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.
- 2024-05-22FDA-DeviceWishbone Medical, Inc.Class IIWishFIX Growth Control Plating System, Model Number: TEFCS28-080US. Orthopedic device.
The products in this lot are incorrectly packed with a Stainless-Steel implant instead of a Titanium implant as the packaging and part number indicated.
- 2024-05-22FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IINamic FLUID DELIVERY SET, REF 91300040
Products have a lack of sterility assurance.
- 2024-05-22FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IInamic CONVENIENCE KIT, a) CAROTID MANIFOLD KIT, REF 60011045; b) 3V MANIFOLD KIT ON WITH NAMIC DT & WASTE MANAGEMENT, REF 60070582; c) LEFT HEART KIT, REF 60071822; d) LEFT HEART KIT, REF 60080085; e) CONVENIENCE KIT, REF 60100055; f) CONVENIENCE KIT, REF 60101041; g) CONVENIENCE KIT, REF 60120336; h) LEFT HEART KIT, REF 60131446; i) 3 VALVE MANIFOLD OFF KIT WITH SQUEEZE CONTRAST CONTROLLER AND PROTECTION STATION, REF 60140798; j) CONVENIENCE KIT, REF 60142592; k) LEFT HEA
Products have a lack of sterility assurance.
- 2024-05-22FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Venous Cannulae: a) DLP¿ 20 Fr. Malleable Single Stage Venous Cannula, Model Number 68120, b) DLP¿ 24 Fr. Malleable Single Stage Venous Cannula, Model Number 68124, c) DLP¿ 28 Fr. Malleable Single Stage Venous Cannula, Model Number 68128, d) DLP¿ 30 Fr. Malleable Single Stage Venous Cannula, Model Number 68130, e) DLP¿ 34 Fr. Malleable Single Stage Venous Cannula, Model Number 68134, f) DLP¿ 12 Fr. Single Stage Venous Cannula, Model Number 67312, g) DLP¿ 16 Fr. Single St
Potential for unsealed sterile packing.
- 2024-05-22FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Suction Tubes: a) DLP¿ Suction Tube 6-Fr. Shaft with Frazier Tip, Model Number: 10050, b) DLP¿ Suction Tube 6-Fr. Shaft with 10-Fr. Soft Tip, Model Number: 10052, c) DLP¿ Suction Tube 6-Fr. Shaft with 10-Fr. Soft Tip, Model Number: 10053, d) DLP¿ Suction Tube 10 Fr. Shaft with 20 Fr. Pool Tip, Model Number: 10060, e) DLP¿ Suction Tube 16 Fr. Shaft with 20 Fr. Fluted Tip, Model Number: 10061
Potential for unsealed sterile packing.
- 2024-05-22FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Intracoronary Shunts: a) ClearView¿ 1.00 mm Intracoronary Shunt, Model Numbers: 31100; b) ClearView¿ 1.25 mm Intracoronary Shunt, Model Numbers: 31125; c) ClearView¿ 2.00 mm Intracoronary Shunt, Model Numbers: 31200
Potential for unsealed sterile packing.
- 2024-05-22FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Disposable Pressure Display Sets: a) DLP¿ 114.3 cm (45 in), Model Number 61000; b) DLP¿ 114.3 cm (45 in), Model Number 62000
Potential for unsealed sterile packing.
- 2024-05-22FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Cardioplegia Adapters: a) DLP¿ 1.8 m (6 ft) Pressure Monitoring Extension Line Adapter, Model Number 25009, b) DLP¿ 1.8 m (6 ft) Pressure Monitoring Extension Line Adapter, Model Number 25010, c) DLP¿ 30.5 cm (12 in) Multiple Perfusion Set, Model Number 14003, d) DLP¿ 38.1 cm (15 in) Multiple Perfusion Set, Model Number 14000, e) DLP¿ 50.8 cm (20 in) Extension Line Adapter, Model Number 11001G, f) DLP¿ Perfusion/Venting Adapter, Model Number 13002, g) DLP¿ Y Adapter
Potential for unsealed sterile packing.
Get notified about new device recalls
Free weekly digest. We email when a new federal recall in this category — or anything else affecting your household — is published.