Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-11-29FDA-DeviceBiomet, Inc.Class IICompress Device Segmental Anchor Plug, 16 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178406
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
- 2023-11-29FDA-DeviceBiomet, Inc.Class IICompress Device Segmental Anchor Plug, 14 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178404
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
- 2023-11-29FDA-DeviceBiomet, Inc.Class IICompress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178402
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
- 2023-11-29FDA-DeviceBiomet, Inc.Class IICompress Device Segmental Anchor Plug, 10 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178400
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIMedline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPRM3312PM, CPRM3312PM2, CPRM3312PMC2, CPRM3312PMTD1, CPRM3312PMX, CPRM3312PWM, CPRM3322FM, CPRM332
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IICuraplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mask, REF # 2442-BVMPCH; 3. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, Infant and Neonate Mask, REF # 2442-BVMCIN; 4. Curaplex Infant Premium BVM with Ba
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN4100MCP; 2. V-Care Infant Resuscitator REF #s: VN3100OB, VN3100MB-2, VN3100MB-M01, VNW3100MB-2, VN3100MBP-3M, VN3100MB, VN3100MB-D1, VN3100MB-M00-0, VN3100MB-M1, VN3100OBP, VN3100MB-B, VN3100MB-I, VN3100MBP-000, VN3100MBP-2, VN3100MB-PW2, VN3100MBP, VN3100MBP-D1, VN3100MBPW-D2, VN3100MBP-F, VN3100MXK-2, VN3100MX-F,
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC STAT-Check manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. STAT-Check Adult Resuscitator REF #: SC9101B; 2. STAT-Check Infant Resuscitator RES #s: SC7101B & SC7101B-M0; 3. STAT-Check Pediatric Resuscitator REF #: SC8121B.
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflatio
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator REF #s: BVMB812, BVMB812-F, BVMB810, BVMB810-F, BVMB810S-F.
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, 9106HZ, 9110HZ, 9120HZ, 9111HZ, 9100HZ, 9105HZ, 9108HZ, 9006HZ, 9002HZ; 2. Horizon Infant Resuscitator REF #s: 6112HZ, 6123HZ, 6125HZ, 6200HZ, 6106HZ, 6118HZ, 6107HZ, 6013HZ, 6117HZ, 6116HZ, 6019HZ, HZI6000, 6009HZ; 3. Horizon Pediatric Resuscitator REF #s: 8106HZ, 8119HZ, 8115HZ, 8107HZ, 8010HZ, 8004HZ, 8105H
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC BreathTech manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. BreathTech Adult Resuscitator REF #s: BT5107F & BT5104P; 2. BreathTech Child Resuscitator REF #s: BT3030, BT3034, & BT3025; 3. BreathTech Infant Resuscitator REF #s: BT2415K, BT2416, BT2032, BT2016, BT2035, BT2216, BT2036, BT2023, BT2026, BT2220-2, BT2020, BT2220, BT2220F, BT2215, BT2415, BT2014, BT2027, BT2015, BT2029, BT2037, BT2048, BT2034, BT2034W, BT2234, BT223
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceSunMed Holdings, LLCClass IIVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6, AF5140MB-F, AF1140MB-T, AF1140MB-PW, AF5140MB-PW, AF1140MB-I, AF5140MB-I, AF5140MBN, AF1140MBS, AF5140MBS, AF1140MB-X, AF5140MB-X, AF1140MB-P, AF5140MBP, AF1140MB-K, AF5140MBS-I,AF5140MBPF-G10, AF5140MBP-G10, AF1140MBP
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- 2023-11-29FDA-DeviceAvanos Medical, Inc.Class IIAVANOS BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI, Product code 221038, provide suction on patients on mechanical ventilation
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.
- 2023-11-29FDA-DeviceInstrumentation LaboratoryClass IIGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
- 2023-11-29FDA-DeviceInstrumentation LaboratoryClass IIGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
- 2023-11-29FDA-DeviceInstrumentation LaboratoryClass IIGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
- 2023-11-29FDA-DeviceSage Products IncClass IIstryker Sage PrimaFit External Urine Management System for the Female Anatomy
A small percentage of devices in three lots of product may contain a tape with a natural rubber latex adhesive which is not declared on the label
- 2023-11-29FDA-DeviceMedtronic NeuromodulationClass IIModel 97745 Controller used with Models 97715 and 97716 Intellis Neurostimulators
Units distributed in Korea and Turkey that did not have the correct firmware installed to support the user interface in the local language.
- 2023-11-29FDA-DeviceBioFire Diagnostics, LLCClass IIFilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, touch screen computer and USB ports. The BioFire FilmArray Torch (BioFire Torch) is an automated in vitro diagnostic (IVD) device intended for use with FDA-cleared or approved IVD BioFire FilmArray Panels. Additional Part/Reference Numbers as part of the system: - HTFA-ASY-0102: BIOFIRE TORCH Duplex. This is a plastic component that covers the BIOFIRE TORCH Modules. The BIOFIRE TORC
Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
- 2023-11-29FDA-DeviceRoche Diagnostics Operations, Inc.Class IITina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311, 501 and 502; INTEGRA 400 plus), 08105618190 - (cobas c 303 and 503)
Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
- 2023-11-29FDA-DeviceBecton Dickinson & CompanyClass IIBD Extension Sets (Product Name, Catalog #) BD SmartSite Filter Extension Set PE lined Standard bore 1 SmartSite bonded 0.2um Large PES Filter 1 PC RLL. L: 16 in PV: 4mL, Catalog # 10013902 BD EXTENSION SET TRI-PORT SMALLBORE TUBING CHECK VALVE 3 SmartSite" VALVES, Catalog # 20062E; BD SECONDARY SET VENTED NONVENTED 20 DROP W/ 2 HANGERS, Catalog # 70000N-07; BD MaxGuardTM tri-fuse extension set, Catalog # ME1224; BD 20-IN EXTENSION SET W/ 0.2 MICRON FILTER ONE SmartSiteTM VALVE AND
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
- 2023-11-29FDA-DeviceBecton Dickinson & CompanyClass IIBD gravity sets (Product Name, Catalog #) 32-IN GRAV SET W/0.2 MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, Catalog # 4030B-07T and BD GRAVITY BLOOD SET 15 DROP 180 MICRON FILTER SmartSiteTM VALVE NONVENTED, Catalog # 42081E.
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
- 2023-11-29FDA-DeviceBecton Dickinson & CompanyClass IIBD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM Pump Infusion Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSiteTM Y-sites, Catalog #
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIUreteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIDuodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IITransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IISmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, and small intestine) by anal insertion. - Model Number: SIF-Q180.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection