- 2021-05-19FDA-DeviceSTEP-HAR MEDICAL LLCClass IAlaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
- 2021-05-19FDA-DeviceROi CPS LLCClass IIRegard Dressing Change Tray
ChloraPrep applicators in the kit can grow organisms that can result in a breach of the applicators package integrity.
- 2021-05-19FDA-DeviceMeridian Bioscience IncClass IIRevogene RNA Software Upgrade Kit Revogene RNA Software Upgrade USBs Catalog Number: 610240 IVD used to install the Revogene System Software version 4.2.0, import the Revogene Instrument Control Protocol file for RNA-based assay, and install the Assay Definition File (ADF) file
Emergency Use Application (EUA) application withdrawn as false negative results reported
- 2021-05-19FDA-DeviceMeridian Bioscience IncClass IIRevogene SARS-CoV-2 Catalog Number: 410700 IVD for real-time Reverse Transcription Polymerase Chain Reaction (rRT-PCR) qualitative test run on the Revogene¿ instrument for the detection of SARS-CoV-2 nucleic acid from nasopharyngeal swab (NPS) specimens from individuals suspected of COVID-19
Emergency Use Application (EUA) application withdrawn as false negative results reported
- 2021-05-19FDA-DeviceFresenius Medical Care Holdings, Inc.Class IIAquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. Fresenius P/N 24-2500-0 All 3 units make up the AquaBplus 2500Fresenius Part Number: G02040109-US AquaBplus 3000 - Typ USA G02040115-US AquaBplus 2500 - Typ USA G02040120FMC-US AquaBplus HF- Typ USA
AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
- 2021-05-19FDA-DeviceSt. Jude Medical, Cardian Rhythm Management DivisionClass IENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
- 2021-05-19FDA-DeviceSt. Jude Medical, Cardian Rhythm Management DivisionClass IASSURITY Pulse Generator REF PM**** SN ********* ST. JUDE MEDICAL Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
- 2021-05-12FDA-DeviceMedtronic NeuromodulationClass IIIntellis Model 97755 Recharger - Product Usage: designed to charge the following Medtronic rechargeable neurostimulators: Intellis with AdaptiveStim Technology Model 97715 Implanted Neurostimulator Intellis Model 97716 Implanted Neurostimulator.
There is a potential for unintended heating that can occur with the Intellis Model 97755 Recharger.
- 2021-05-12FDA-DeviceGreiner Bio-One North America, Inc.Class IIVACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing.
VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays.
- 2021-05-12FDA-DeviceGE Healthcare, LLCClass IICentricity Universal Viewer - Product Usage: intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.
When the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s comparison exam(s).
- 2021-05-12FDA-DeviceMedtronic Inc.Class IIMedtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction.
Incorrect size printed on the device; packaging is labeled correctly.
- 2021-05-12FDA-DeviceLeMaitre Vascular, Inc.Class IILeverEdge Contrast Injector: 35 cc syringe with ISO 594 female luer lock - Product Usage: intended to be used to perfuse contrast medias into the vessels for angiographic procedures, Catalog #: 4100-00.
The firm received a complaint regarding particulate found in the packaging foam that could contaminate the sterile field.
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IICareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIProgrammers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIImplantable cardiac resynchronization therapy pacemakers (CRT-P) - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Percepta family: Percepta CRT-P MRI SureScan, Percepta Quad CRT-P MRI SureScan. 2. Serena family: Serena CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan. 3. Solara family: Solara CRT-P MRI SureScan, Solara Quad CRT-P MRI SureScan.
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIImplantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR MRI SureScan. 2. Astra family: Astra XT DR MRI SureScan, Astra XT SR MRI SureScan.
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
- 2021-05-12FDA-DeviceEncore Medical, LPClass IIDJO EMPOWR KNEE Femoral Sizer
Incorrect assembly of a subcomponent by a vendor and as a result, the windows provided to show the size are facing the opposite direction.
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IICRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
- 2021-05-12FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IICareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
- 2021-05-12FDA-DeviceOto Med IncClass IIGrace Medical REF S-4500 *** spineX BONE DUST COLLECTOR *** 2 stainless steel filters - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.
- 2021-05-12FDA-DeviceOto Med IncClass IIGrace Medical REF S-3500 *** COX BONE DUST COLLECTOR *** 2 stainless steel filters - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.
- 2021-05-12FDA-DeviceOto Med IncClass IIGrace Medical REF S-2500 *** SHEEHY BONE DUST COLLECTOR *** Stainless steel filter - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.
- 2021-05-05FDA-DeviceOmnia SRLClass IIOmnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32
Omnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.
- 2021-05-05FDA-DevicePhilips North America LlcClass IIPhilips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring Service numbers associated with FC2010: Xper Flex Cardio 2010 Rev A 453564241901 Xper Flex Cardio 2010 Rev B 453564483321 Flex Cardio FC2010 Rev C 453564621791 Xper Flex Cardio 2010 REV C RUEXCHANGE 453564675021 Xper Flex Cardio 2010 Rev D 453564669081 Xper Flex Cardio 2010 Rev D Russia 453564674581
Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
- 2021-05-05FDA-DeviceSmiths Medical ASD Inc.Class IIACE MDI Spacer with Small Mask - Product Usage: designed to assist with the delivery of aerosolized medications from metered dose inhaler canisters and with mechanically ventilated patients.
The firm became aware via complaints that one lot of ACE MDI Spacers with Small Masks and one lot of Palm Cups (Large) may have incorrect shipping box labels.
- 2021-05-05FDA-DeviceSmiths Medical ASD Inc.Class IIPalm Cups - Product Usage: intended to be used to percuss the chest wall to mobilize the secretions within the bronchial tree.
The firm became aware via complaints that one lot of ACE MDI Spacers with Small Masks and one lot of Palm Cups (Large) may have incorrect shipping box labels.
- 2021-05-05FDA-DeviceDJ Orthopedics de Mexico S.A. de C.V.Class IIDONJOY REF:11-0473-9 ULTRASLING PRO AB, UNIVERSAL, Rx ONLY, UDI: (01)00190446683681 - Product Usage: intended for medical purposes to immobilize the arm, by means of a fabric band suspended from around the neck.
failed rivet at the chafe tab of arm sling.
- 2021-05-05FDA-DeviceDJ Orthopedics de Mexico S.A. de C.V.Class IIDONJOY REF:11-0472-9 ULTRASLING PRO ER/IR 30, UNIVERSAL, Rx ONLY, UDI: (01)00190446683674 - Product Usage: intended for medical purposes to immobilize the arm, by means of a fabric band suspended from around the neck.
failed rivet at the chafe tab of arm sling.
- 2021-05-05FDA-DeviceDJ Orthopedics de Mexico S.A. de C.V.Class IIDONJOY REF:11-0471-9 ULTRASLING PRO ER/IR 15, UNIVERSAL, Rx ONLY, UDI: (01) 00190446683667 - Product Usage: intended for medical purposes to immobilize the arm, by means of a fabric band suspended from around the neck.
failed rivet at the chafe tab of arm sling.
- 2021-05-05FDA-DeviceDJ Orthopedics de Mexico S.A. de C.V.Class IIDONJOY REF:11-0447-9-ULTRASLING PRO, UNIVERSAL, Rx ONLY, UDI:(01)00190446673552 - Product Usage: intended for medical purposes to immobilize the arm, by means of a fabric band suspended from around the neck.
failed rivet at the chafe tab of arm sling.