- 2026-07-22FDA-DeviceOrthofixClass IIOrthofix Minirail external fixation devices recalled for incorrect labeling
Orthofix Minirail orthopedic external fixation devices were shipped with incorrect model labels. Some M101 devices were labeled as M103, M103 devices were labeled as M101, and M122 devices were labeled as M111. This labeling error could cause confusion during surgical planning and device selection.
- 2026-07-22FDA-DeviceGEClass IINationwideGE Optima XR radiography systems recalled for arm drop risk
GE HealthCare identified that some Optima XR200/220/240amx systems that underwent column replacement may not have had proper safety testing completed. If an internal cable fails, the safety mechanism could malfunction, allowing the X-ray arm to drop unexpectedly during patient exams.
- 2026-07-22FDA-DeviceBoston ScientificClass IINationwideBoston Scientific injection system recalled for syringe disconnection risk
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the device manifold. This could result in fluid leakage, inability to inject or aspirate properly, or complete separation of the syringe, potentially requiring device replacement and extending the procedure.
- 2026-07-22FDA-DeviceICU MedicalClass IINationwideICU Medical Plum Solo IV pump recalled for overdose risk
In certain conditions, the Plum Solo Precision IV pump (v1.0.3) may not properly track the volume of fluid being delivered and could administer up to twice the intended amount of medication or fluid. This affects specific serial numbers of the pump.
- 2026-07-22FDA-DeviceICU MedicalClass IINationwideICU Medical Plum Duo infusion pump recalled for overdose risk
Certain ICU Medical Plum Duo infusion pumps may fail to track the amount of fluid delivered on the screen display. In rare cases, this could cause the pump to deliver up to twice the intended volume of medication or fluid. This affects specific serial numbers of two item numbers (400020401 and 400020403) running software version 1.2.3.
- 2026-07-22FDA-DeviceIntuitive SurgicalClass IINationwideIntuitive Surgical da Vinci drape recalled for detachment risk
Arm drapes used with Intuitive Surgical's da Vinci SP robotic surgical system may detach from the instrument arm due to a defect in the drape rings at the top of the product. If a drape detaches during surgery, it can expose part of the instrument arm and may delay the procedure or require switching to an alternative surgical approach.
- 2026-07-22FDA-DeviceOlympusClass IINationwideOlympus Ellik Evacuator glass body recalled for sterilization issue
Olympus has recalled the Glass Body for the 194 Ellik Evacuator (Model 190A) because some sterilization instructions in the user manual may not reliably make the device sterile. The affected sterilization parameters should no longer be used.
- 2026-07-22FDA-DeviceOlympusClass IINationwideOlympus Ellik Evacuator recalled for sterilization risk
Olympus has recalled the Ellik Evacuator (Model 194), a medical device used during urological procedures. Some sterilization methods described in the device instructions may not reliably sterilize the device, creating a potential infection risk.
- 2026-07-22FDA-DeviceOlympusClass IINationwideOlympus Toomey Evacuator 50cc recalled for sterilization risk
Olympus has recalled the Toomey Evacuator 50cc with Adaptor (Model 188A) because some sterilization instructions in the product manual may not reliably make the device fully sterile. Affected units were distributed worldwide, including across the US, Canada, Japan, Europe, and Latin America.
- 2026-07-22FDA-DeviceOlympusClass IINationwideOlympus Toomey Evacuator recalled for sterilization risk
Olympus has recalled the Toomey Evacuator 30cc with Adaptor (Model 188) because some of the sterilization instructions in the device manual may not reliably make the device sterile. The affected instructions should no longer be used.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical bone marrow tray recalled for sterility failure
Argon Medical Devices is recalling bone marrow tray kits (models CB0033TL and CB0355TLA) that contain lidocaine injections. The lidocaine ampules in these kits did not meet sterility standards required for safe use. The kits were distributed nationwide and internationally.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical Arterial Line Kit recalled for sterility failure
Argon Medical Devices is recalling certain Arterial Line Kits (model 20ga x 6, REF: 400115A) that contain Lidocaine HCL Injection. These kits failed sterility testing and do not meet safety release criteria. The affected kits have specific lot numbers with expiration dates in July 2027.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical T-Lok Bone Marrow Tray recalled for sterility failure
Argon Medical Devices is recalling T-Lok Bone Marrow Tray kits that contain Lidocaine HCL Injection ampules. These kits failed to meet sterility standards during quality testing. The affected kits were distributed nationwide and internationally.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical biopsy kit recalled for sterility failure
Argon Medical Devices is recalling General Biopsy Trays (REF: GUTS1000) that contain Lidocaine HCL Injection ampules. The lidocaine ampules in these kits did not meet sterility release criteria, meaning they may not be safe for use. Approximately 1,800 kits were distributed nationwide and to Trinidad and Tobago.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical introducer trays recalled for sterility issue
Argon Medical Devices is recalling One-Piece Introducer Trays (REF 497455 and 497434) that contain Lidocaine HCL Injection ampules. The Lidocaine ampules in these kits did not meet sterility release criteria, meaning they were not verified as free from contamination before distribution.
- 2026-07-22FDA-DeviceArgon Medical DevicesClass INationwideArgon Medical pacing electrode trays recalled for sterility risk
Argon Medical Devices is recalling certain transvenous pacing electrode trays (model 008416A and 008566A) because they contain lidocaine ampules that did not meet sterility standards. The affected kits were distributed nationwide and internationally. No injuries have been reported.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Intro Kit introducer recalled for leakage risk
Abiomed heart pump introducer sheaths in the Intro Kit may leak fluid from under the sheath cap or along the hub score lines. The manufacturing defect is not visible to users. All current batches of the kit worldwide are affected.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella CP heart pump recalled for leakage risk
Abiomed has recalled the Impella CP heart pump due to a potential for fluid leakage from the introducer sheath (a tube inserted during the procedure). The leakage can occur under the sheath cap or along connection points, and these defects cannot be seen by the user. All current batches of the pump set are affected worldwide.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella CP heart pump recalled for sheath leakage risk
Abiomed has recalled its Impella CP Smart Assist Set heart pump due to a potential leak in the introducer sheath (a thin tube used during insertion). The leak can occur under the sheath cap or along seam lines in certain models, but the defect is not visible to the user. All current batches worldwide are affected.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella CP heart pump recalled for sheath leakage risk
Abiomed has recalled the Impella CP heart pump set due to a manufacturing defect in the introducer sheath that can cause leakage. The leakage can occur under the sheath cap or along hub score lines in the 14Fr and 23Fr introducers included with every pump set, and the defect is not visible to users.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella 5.5 heart pump recalled for sheath leakage risk
Abiomed has recalled certain Impella 5.5 SmartAssist S2 heart pump sets worldwide due to a manufacturing defect in the introducer sheaths. The 14Fr and 23Fr introducers included in these pump sets may leak fluid from under the sheath cap or along hub seams, and the leakage cannot be seen by the user before or during use.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella 5.5 heart pump recalled for leakage risk
The Abiomed Impella 5.5 Smart Assist Set heart pump contains an introducer sheath that may leak under the cap or along hub seams. This manufacturing defect is not visible to the user but is present in all affected pump sets. The issue affects 383 units distributed worldwide.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Introducer Kit recalled for sheath leakage risk
Abiomed has recalled certain Introducer Kit 23 Fr units (Pump Set Code 1000441) due to a potential for sheath leakage from under the sheath cap and along hub score lines. This manufacturing issue cannot be seen by the user and affects introducers included in every pump set.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella CP Introducer Kit recalled for leakage risk
Abiomed has recalled the Impella CP Introducer Kit (Pump Set Code 1000550) worldwide due to a potential for fluid leakage from under the sheath cap and along hub score lines in the 14Fr and 23Fr introducers. The leakage defect is not visible to the user and affects all current batches of this heart pump accessory kit.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella CP introducer recalled for leakage risk
Abiomed has recalled the Impella CP heart pump introducer due to a manufacturing defect that can cause fluid leakage from under the sheath cap and along the hub. The defect is not visible to the user and affects all current batches worldwide.
- 2026-07-22FDA-DeviceAbiomedClass INationwideAbiomed Impella RP Flex recalled for sheath leakage risk
The Abiomed Impella RP Flex heart pump set includes introducers that may leak fluid from under the sheath cap or along hub score lines due to manufacturing defects. These leaks cannot be seen by visual inspection. All current batches worldwide are affected.
- 2026-07-22FDA-FoodAmy's KitchenClass IIAmy's Kitchen lentil soup recalled for spoilage risk
Amy's Kitchen organic lentil soup (light in sodium, 14.5 oz cans) with lot code 60D0924 and best-before date of April 2027 may spoil before the printed date. The affected product was distributed across the U.S. and Canada.
- 2026-07-22FDA-FoodNutriciaClass IINutricia NOURISH meal replacement formula recalled for low vitamin A
Nutricia North America discovered that its testing laboratory used an incorrect method to measure vitamin A content in NOURISH ORGANIC WHOLE FOODS MEAL REPLACEMENT PEPTIDE FORMULA BERRY MEDLEY. The test results showed higher vitamin A levels than what is actually in the product, meaning affected packages contain less vitamin A than the label states.
- 2026-07-22FDA-FoodNutriciaClass IINutricia Liquid Hope formula recalled for low Vitamin A
Nutricia North America discovered that its testing lab used an incorrect method to measure Vitamin A in Liquid Hope Organic Whole Foods Meal Replacement Peptide Formula (Berry Medley). As a result, the product contains less Vitamin A than the label states. The affected pouches have best-if-used-by dates through March 15, 2028 (24-count cases) and October 7, 2027 (12-count packages).
- 2026-07-22FDA-DrugCareFusionClass IINationwideBD ChloraPrep Clear recalled for packaging seal defect
BD ChloraPrep Clear, a sterile skin antiseptic solution used before medical procedures, may have open or incomplete seals on individual applicator packaging. This could allow contamination of the sterile product. The issue affects multiple lot numbers distributed nationwide.
Get notified about new Colorado recalls
Free weekly digest. Add Colorado to your profile to filter alerts to your state.
Get the Sunday Brief