Quantum Rehab Q6 Edge HD wheelchair recalled for possible fall risk
Quantum Rehab Q6 Edge HD powered wheelchairs with controller software version 2.0.11 may show reminder pop-ups if the "Remind Me Later" option is selected. While driving, this may cause an unexpected park brake stop, which may result in a fall or injury. While not driving, the control system may crash and remain in pre-operational mode. No injuries have been reported.
From the FDA notice · Read the official notice ↗
What to do
- Check whether your Q6 Edge HD controller runs software version 2.0.11.
- Contact Pride Mobility Products Corp. about the update to software version 2.0.13. The notice does not say how to get it.
- The notice does not give a remedy. Ask Pride Mobility Products Corp. or your equipment provider how to proceed.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 10 products — see all of them.
Details
- Units affected
- 1102
- Illnesses reported
- None reported
- Manufactured in
- 401 York Ave, Duryea, PA, United States
- Classification
- Class II
- Recall number
- Z-2845-2026
In the FDA's words
- Description
- Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. Lot Code: Q6 Edge HD 0060650920005*9 K042612 D049086 41875 ITI Distribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia. Affected quantity: 1102
- Hazard
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
10 products in this recall
Quantum Rehab power wheelchairs recalled for unexpected brake stop risk
Quantum Rehab is recalling 10 powered wheelchair models and controller products, including Quantum 4Front 2, Quantum Outback, Q6 Edge 3 Stretto and Quantum Rival. If the "Remind Me Later" option is selected for user reminders, reminder pop-ups may appear during use. While driving, this may cause an unexpected park brake stop that could result in a fall or injury; while not driving, the control system may crash and stay in pre-operational mode.
The agency lists each product separately. Check every one you might have.
- Brand Name: Quantum Rehab Product Name: Quantum 4Front 2 Model/Catalog Number: Quantum 4Front 2 Software Version: 2.0.11, new version: 2.0.13 Product Des…Z-2841-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Quantum Outback Model/Catalog Number: Quantum Outback Software Version: 2.0.11, new version: 2.0.13 Product Descr…Z-2842-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product D…Z-2843-2026 · 1102
- Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair…Z-2844-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wh… (this page)Z-2845-2026 · 1102
- Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description:…Z-2846-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description…Z-2847-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Whee…Z-2849-2026 · 1102
- Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair…Z-2852-2026 · 1102
- Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Descripti…Z-2853-2026 · 1102