Serious riskFDA DeviceClass II

Quantum Rehab 4Front wheelchair software recalled for possible falls

Quantum Rehab 4Front powered wheelchair controller software version 2.0.11 may show reminder pop-ups if the "Remind Me Later" option is selected. While the wheelchair is being driven, this may cause an unexpected park brake stop, which may result in a fall or injury. While the wheelchair is not being driven, the control system may crash and stay in pre-operational mode. No injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Do not select the "Remind Me Later" option for user reminders.
  2. Check whether your 4Front controller runs software version 2.0.11. The recall covers this version; the new version is 2.0.13.
  3. Contact Pride Mobility Products Corp. or Quantum Rehab for instructions, since the notice does not specify a remedy.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 10 products — see all of them.

Details

Units affected
1102
Illnesses reported
None reported
Manufactured in
401 York Ave, Duryea, PA, United States
Classification
Class II
Recall number
Z-2852-2026

In the FDA's words

Description
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. Lot Code: 4Front 0060650920003*5 K143383 D246699 41875 ITI Distribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia. Affected quantity: 1102
Hazard
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.

10 products in this recall

Quantum Rehab power wheelchairs recalled for unexpected brake stop risk

Quantum Rehab is recalling 10 powered wheelchair models and controller products, including Quantum 4Front 2, Quantum Outback, Q6 Edge 3 Stretto and Quantum Rival. If the "Remind Me Later" option is selected for user reminders, reminder pop-ups may appear during use. While driving, this may cause an unexpected park brake stop that could result in a fall or injury; while not driving, the control system may crash and stay in pre-operational mode.

The agency lists each product separately. Check every one you might have.