- 2021-01-27FDA-DevicePenumbra Inc.Class IPenumbra JET 7 Reperfusion Catheter REF 5MAXJET7 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- 2021-01-20FDA-DevicePhilips Ultrasound IncClass IIAll EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIQ CVx, Affiniti 30, Affiniti 50, Affiniti 70. Diagnostic Ultrasound System. MOD - EPIQ 5 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 7 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 100-240V 50/60 Hz, 600 VA MOD - Affinity 30 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 50 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 70 GMDN - 40761 100-240V 50/60
The manufacturer has determined that with certain uncommon workflows there is potential for incorrect patient data to be displayed and saved into an exam. The issue(s) manifest differently for different versions of software.
- 2021-01-20FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM 1600 Analyzer - In vitro diagnostic testing of clinical specimens. Product Siemens Material Number (SMN): 11066000 - Product Usage: intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.
Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
- 2021-01-20FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM 1300 Analyzer - In vitro diagnostic testing of clinical specimens. Product Siemens Material Number (SMN): 11066001 - Product Usage: intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.
Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
- 2021-01-20FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica CH 930 Analyzer - In vitro diagnostic testing of clinical specimens. Product Siemens Material Number (SMN): 11067000 - Product Usage: intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.
Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
- 2021-01-20FDA-DeviceMedtronic NeuromodulationClass IIMedtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.
A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.
- 2021-01-20FDA-DeviceICU Medical, Inc.Class IIicumedical Cogent" Hemodynamic Monitoring System - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician.
Due to a potential software issue, the display may show the incorrect continuous cardiac output (CCO) values after PulseCO calibration.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIOptima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIDiscovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, usin
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIDiscovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIGE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIGE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DeviceGE Healthcare, LLCClass IIGE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- 2021-01-20FDA-DevicePhilips North America, LLCClass IIEmission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
- 2021-01-20FDA-DeviceVISBY MEDICAL INCClass IIvisby medical COVID-19 Test kit *** An in vitro diagnostic for the qualitative detection of SARS-CoV-2 - Product Usage: The Visby Medical COVID-19 Test is only for use under the Food and Drug Administration s Emergency Use Authorization.
Two issues: 1) 5 Lots were found to have incorrect labeling on the outer box stating that the product is CLIA Waived. 2) 3 Lots contain a kit component that expires prior to the expiration date on the test kits.
- 2021-01-20FDA-DeviceRoche Molecular Systems, Inc.Class IIThe cobas z 480 analyzer including dedicated software for IVD use is a plate-based automated amplification and detection system for nucleic acids. The analyzer is intended to be used as an in-vitro diagnostic or screening system providing real time PCR amplification and detection for human samples within the system. The cobas x 480 instrument and cobas z 480 analyzer are to be used by laboratory professionals trained in laboratory techniques and by instruction on the use of the system and ana
Dirty Lens May Cause Invalid or False Positive Results
- 2021-01-20FDA-DeviceOrtho Clinical Diagnostics IncClass IIVITROS XT Chemistry Products ALB-TP Slides (ALB is not affected) - IVD quantitatively measures total protein (TP) concentration in serum and plasma using VITROS XT 3400 Chemistry Systems and VITROS XT 7600 Integrated Systems Product Code: 6844293
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides
- 2021-01-20FDA-DeviceOrtho Clinical Diagnostics IncClass IIVITROS Chemistry Products TP Slides- IVD measure total protein (TP) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems Product Code: 8392292
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides
- 2021-01-20FDA-DeviceMedtronic Inc.Class IICareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes Software Uploader versions: 3.1.5.000 3.1.2.000
Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software.
- 2021-01-20FDA-DeviceAnsh Labs, LLCClass IIAnshLabs - SARS-CoV2 IgM ELISA (u-Capture), IVD, REF: AL-1002, kit contents, Ab PLATE 1 X ! Plate, IgM Sample DIL 1 x 100 mL, ENZ CONJ CONC 1 x 1.5 mL, Calibrators A-C 3 x 1.2 mL, CONJ DIL 1 X 12 mL, Stop SOLN 1 X 12 mL, TMB SOLN 1 X 12 mL, WASH CONC A 1 X 60 mL, (01)00853592008219 - Product Usage: intended for the qualitative detection of IgM antibodies in serum collected from the individuals suspected with signs and symptoms of COVID-19 infection by their healthcare provider.
Labeling update; To reduce the risk of false positive/false negative results the IFU for the adaptive immune response to SRS-CoV-2 testing is being updated to remove Plasma as a "sample type" and to update the "limitation" section to include potential cross-reactivity with non-SARS-CoV2 strains.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IICitrasate Liquid Acid concentrate drums, Product Code 13-2251 -CA - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate drums, Product Code 13-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate drums, Product Code 13-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-4231-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-2351-8 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IICitrasate Liquid Acid concentrate bottles, Product Code 08-2251-CA - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- 2021-01-20FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.