- 2021-02-03FDA-DeviceRespironics California, LLCClass IIPhilips Respironics V60 Ventilator Part Number 1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
- 2021-02-03FDA-DeviceRespironics California, LLCClass IIPhilips Respironics V60 Ventilator Part Number 1053616 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
- 2021-02-03FDA-DeviceRespironics California, LLCClass IIPhilips Respironics V60 Ventilator Part Number 1053615 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
- 2021-02-03FDA-DeviceRespironics California, LLCClass IIPhilips Respironics V60 Ventilator Part Number 1053614 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
- 2021-02-03FDA-DeviceRespironics California, LLCClass IIPhilips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
- 2021-02-03FDA-DeviceNihon Kohden America IncClass IIPSG-1100 Sleep Diagnostic System. Sample product label: JB-110A SERIAL No. 01454 RoHS Compliant; PSG-100 Head Box; UDI (01) 04931921108383; Made in Japan The PSG-1100 Sleep Diagnostic System is a digital PSG amplifier intended to record the electrical activity of the brain (EEG) and other bio-potential signals and to record physiological data required for sleep studies.
Medical device non-conformance to electrical safety standard (IEC 60601-1).
- 2021-02-03FDA-DevicePHILIPS HOME HEALTHCARE SOLUTIONClass IIProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
- 2021-02-03FDA-DeviceEnztec LimitedClass IIREF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.
- 2021-01-27FDA-DeviceExactech, Inc.Class IIEquinoxe Humeral Stem, Primary, Press Fit, 11mm
Potential for incorrect Instructions For Use (IFU) included with an Equinoxe Humeral Stem, Primary, Press Fit, 11mm package.
- 2021-01-27FDA-DeviceRoche Diagnostics Operations, Inc.Class IIA calculator/data processing module for clinical use is an electronic device intended to store, retrieve, and process laboratory data Catalog Number: 07154003001
A software error results in the unintentional removal of the serum-indices flag that would otherwise prevent the release of results. The Cobas Inifinity has an auto-verification feature to hold results for manual review when they meet specific criteria. The serum indices flag is used for serum indices tests that are performed to assess the quality of the sample (e.g. hemolysis, icterus, and lipemia). Normally, when the Cobas Infinity receives a result for a test that is serum indices-sensitive, it flags the result and the software then waits for the results of the serum indices tests before validating or rejecting the test result. Roche has discovered a software error that under specific conditions causes the flag to be incorrectly removed. This allows for the possibility that a sample of poor quality may return an unreliable/incorrect test result that is mistakenly reported to the health care provider and/or patient without the proper disclaimer that the result is based on a sample of poor quality.
- 2021-01-27FDA-DeviceBiomet, Inc.Class IIA.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.
The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
- 2021-01-27FDA-DeviceAbbott Point Of Care Inc.Class IIAbbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood. List Number: 03P76-25
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
- 2021-01-27FDA-DeviceAbbott Point Of Care Inc.Class IIAbbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
- 2021-01-27FDA-DeviceBoston Scientific CorporationClass IEMBLEM S-ICD Subcutaneous Electrode Model 3501. Component of implantable cardioverter defibrillator (ICD)
There is potential for fractures which results in the inability to delivery therapy.
- 2021-01-27FDA-DeviceFujifilm Medical Systems U.S.A., Inc.Class IISynapse PACS Software Versions 5.1 and higher
There is a potential for the wrong patient information may be displayed in the viewer or PowerJacket. Additionally, the incorrect max standard uptake values (SUV) for PET ad CT studies may be returned for Siemens Modality.
- 2021-01-27FDA-DeviceRoche Molecular Systems, Inc.Class IIuPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001
When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.
- 2021-01-27FDA-DeviceSteris CorporationClass IIReliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.
- 2021-01-27FDA-DeviceBeckman Coulter Inc.Class IIACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 - Product Usage: use on the Access Family of Immunoassay Systems only.
The Quality Control (QC) card has a label error for the standard deviation (SD) values for one out of the two levels of QC.
- 2021-01-27FDA-DeviceSmiths Medical ASD Inc.Class IIPortex Acapella DH Vibratory PEP Device - Product Usage: a single patient use device that provides Positive Expiratory Pressure (PEP) Therapy for patients who have Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems, and patients with atelectasis.
Shipping box may contain wrong model.
- 2021-01-27FDA-DeviceMedtronic NeurosurgeryClass IIREF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.
Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction
- 2021-01-27FDA-DeviceMedtronic NeurosurgeryClass IIREF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.
Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction
- 2021-01-27FDA-DeviceMedtronic NeurosurgeryClass IIREF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.
Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction
- 2021-01-27FDA-DeviceBoston Scientific CorporationClass IIACUITY Universal Cutter, MODEL 7060 - Product Usage: intended to facilitate guide catheter removal after the Boston Scientific coronary venous lead is positioned.
Boston Scientific (BSC) is retrieving five Model 7060 ACUITY cutters distributed in the US that did not undergo the applicable sterile processing procedure.
- 2021-01-27FDA-DeviceThe Magstim Company LimitedClass IINeurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.
A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.
- 2021-01-27FDA-DeviceLaerdal Medical (Suzhou) Co., Ltd.Class IILaerdal Compact Suction Unit (LCSU 4), Catalog Numbers 880051, 880061, 881151, 880052, 880062, 881152 - Product Usage: It is intended for intermittent operation to remove secretions, blood or vomit from a patient s airway to allow ventilation.
Component failures may cause the unit to lose all suction. Users will need to obtain alternate suction devices or methods, leading to a delay in treatment and, in worst case scenario, potential suffocation if alternative suction or ventilation method is not administered by the user.
- 2021-01-27FDA-DeviceSkeletal Kinetics, LlcClass IIOsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
- 2021-01-27FDA-DeviceSkeletal Kinetics, LlcClass IIOsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
- 2021-01-27FDA-DevicePenumbra Inc.Class IPenumbra JET 7MAX KIT Penumbra JET 7 Reperfusion Catheter+ MAX Delivery Device + Hi-Flow Aspiration Tubing REF J7XKIT132 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration System
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- 2021-01-27FDA-DevicePenumbra Inc.Class IPenumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7L138KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- 2021-01-27FDA-DevicePenumbra Inc.Class IPenumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.