- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips X3-1 ultrasound transducer labeling
Philips is clarifying labeling on certain X3-1 ultrasound transducers to better define their useful lifespan in clinical use. This affects a small number of devices distributed nationwide. No injuries have been reported.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips S5-2 ultrasound transducer labeling
Philips Ultrasound, Inc. is providing clarification and updated labeling for the S5-2 Ultrasound Transducer to define its useful life in the field. This affects a limited number of units distributed nationwide.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck your Philips S4-1 transducer labeling
Philips is recalling certain S4-1 ultrasound transducers to clarify and update labeling that defines how long these devices can be safely used in clinical settings. No injuries have been reported. The recall affects 41 units distributed nationwide.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips OMNI III TEE transducer expiration
Philips is clarifying and updating labeling on certain OMNI III TEE ultrasound transducers to better define their useful life in the field. This is a precautionary measure affecting 75 units distributed nationwide.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips OMNI II TEE transducer useful life label
Philips is clarifying the useful life labeling on certain OMNI II TEE Ultrasound Transducers to ensure healthcare facilities have accurate information about when these devices should be replaced. Two units with specific serial numbers (02P8MV and 02MH3J) were affected and distributed nationwide.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips ultrasound transducer useful life labeling
Philips is clarifying labeling on certain Mini Multi TEE Ultrasound Transducers to better define when these devices should no longer be used in clinical settings. The affected units may lack clear guidance on their useful lifespan, which could affect safe operation.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips L17-5 ultrasound transducer labeling
Philips is clarifying labeling on certain L17-5 Ultrasound Transducers to better define their useful life in clinical use. This affects 2 units with specific model and serial numbers distributed nationwide. No injuries have been reported.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips L12-5 ultrasound transducer labeling
Philips is clarifying and updating labeling on certain L12-5 ultrasound transducers to better define their useful life in the field. This affects 40 units distributed nationwide. No injuries have been reported.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck your Philips C9-4 ultrasound transducer
Philips Ultrasound, Inc. is clarifying labeling on the C9-4 ultrasound transducer to better define how long the device remains safe and effective in clinical use. This affects a very limited number of units distributed nationwide.
- 2026-04-01FDA-DevicePhilipsClass IIINationwideCheck Philips 3D6-2 transducer useful life labeling
Philips is clarifying and updating labeling on certain 3D6-2 Ultrasound Transducers to better define their useful life in the field. This affects a limited number of units distributed nationwide. No injuries have been reported.
- 2026-04-01FDA-DevicePhilipsClass INationwideCheck your Philips Respironics ventilator settings
Philips Respironics has recalled Trilogy Evo Universal ventilators (software version 1.05.15.00) because using certain non-pneumatic nebulizers with these machines may cause a mismatch between the breathing volume set by your doctor and the volume your lungs actually receive. This could affect the effectiveness of your ventilation therapy.
- 2026-04-01FDA-DevicePhilipsClass INationwideStop using Philips Respironics Trilogy EV300 with non-pneumatic nebulizers
Philips Respironics has recalled Trilogy EV300 ventilators (software version 1.05.15.00) because using them with certain types of nebulizers can cause the device to deliver a different amount of air to the patient than what was set. This mismatch between the intended and actual breath volume could affect patient safety.
- 2026-04-01FDA-DevicePhilipsClass INationwideCheck your Philips Respironics ventilator settings
Philips Respironics has recalled certain Trilogy Evo O2 home ventilators because using non-pneumatic nebulizers with them may cause a mismatch between the breath volume set on the machine and the volume actually delivered to the patient. This discrepancy could result in the patient receiving too much or too little air during ventilation.
- 2026-04-01FDA-DevicePhilipsClass INationwideCheck your Philips Respironics Trilogy Evo ventilator
Philips Respironics has identified an issue with Trilogy Evo ventilators when used with certain non-pneumatic nebulizers. The ventilator may deliver a different amount of air to the patient than the setting shows, which could affect treatment effectiveness. Philips is working with health care providers and regulators on a solution.
- 2026-04-01FDA-DeviceOlympusClass IINationwideCheck your Olympus endoscope reprocessor compatibility
Olympus has identified that two adapter models (MAJ-1443 and MAJ-1444) are no longer compatible with OER-Pro and OER-Elite automated endoscope reprocessors. This means these adapters cannot be safely used with these reprocessing machines. No injuries have been reported.
- 2026-04-01FDA-DeviceOlympusClass IINationwideCheck your Olympus endoscope reprocessor compatibility
Olympus has identified that certain adapter models (MAJ-1443 and MAJ-1444) are no longer compatible with the OER-Elite and OER-Pro automated endoscope reprocessors. This incompatibility could affect proper cleaning and sterilization of medical equipment.
- 2026-04-01FDA-DeviceOlympusClass IINationwideStop using Olympus MAJ-1444 valve with OER reprocessors
Olympus endoscope air/water valves (models MAJ-1443 and MAJ-1444) are no longer compatible with OER-Pro and OER-Elite automated endoscope reprocessors. Using these valves with these reprocessors may result in incomplete or improper cleaning of endoscopes.
- 2026-04-01FDA-DeviceOlympusClass IINationwideStop using Olympus MAJ-1443 suction valve with OER reprocessors
Olympus endoscope suction valves (models MAJ-1443 and MAJ-1444) are no longer compatible with Olympus OER-Pro and OER-Elite automated cleaning systems. Using these valves with these reprocessors may result in improper cleaning and sterilization of medical equipment.
- 2026-04-01FDA-DeviceBeckman CoulterClass IINationwideStop using Beckman Coulter MicroScan Neg MIC 56
Beckman Coulter has recalled MicroScan Neg MIC 56 antimicrobial susceptibility testing panels (Catalog C42464, Lot 2026-01-03) due to potential contamination in the test wells. These panels are used by laboratories to identify bacteria and test antibiotic resistance. Contaminated panels could produce inaccurate test results.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI safety if you have I.T.S. pelvic implant
I.T.S. GmbH has recalled certain plates and screws used in pelvic reconstruction surgery because new testing shows they may heat up more than expected during MRI scans under certain conditions. This was not fully described in the original instructions. No injuries have been reported.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck if your implant needs MRI precautions
I.T.S. GmbH has recalled certain cortical screws, locking stabilization screws, and HLS plates used in hand surgery. New MRI safety testing found these implants may heat up more than expected during MRI scans under certain conditions. The original instructions did not reflect this updated safety information.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck your surgical implant lot number
I.T.S. GmbH headless compression screws used in orthopedic surgery may heat up more than expected during MRI scans under certain conditions. Updated safety testing found higher temperature increases than the original instructions stated. This affects specific lot numbers of these implants distributed across the United States.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI compatibility of I.T.S. locking plates
I.T.S. GmbH has recalled certain foot locking plates and hallux osteotomy locking plates used in orthopedic surgery. New MRI safety testing found that these implants may heat up more than expected during MRI scans under certain conditions, which was not fully described in the original instructions.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI safety for I.T.S. ulna bone implants
I.T.S. GmbH is recalling specific DUL and UOL bone plate and screw systems used to stabilize ulna fractures. New MRI safety testing found these implants generate more heat during MRI scans than the original instructions stated, which could potentially cause tissue damage.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideContact your surgeon about I.T.S. orthopedic plates
I.T.S. GmbH recalled certain distal femur and tibia locking plates used in orthopedic surgery after updated MRI safety testing found they produce higher temperature increases during MRI scans than the original instructions stated. This could potentially cause tissue heating at the implant site during an MRI procedure.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI safety if you have I.T.S. humeral plate
I.T.S. GmbH distal humeral plates and locking cortical screws used in arm surgery may heat more than expected during MRI scans under certain conditions. New testing found the heating levels are higher than what the original instructions stated.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI safety if you have I.T.S. fibula plate
I.T.S. GmbH fibula plates used in orthopedic surgery have been recalled because new MRI safety testing found they heat up more than the instructions originally stated when exposed to certain MRI machine conditions. This could potentially cause tissue burns during an MRI scan.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck I.T.S. GmbH pelvic implants before MRI
I.T.S. GmbH has recalled certain Pelvic Reconstruction System components (Fixation Screws and Spike short devices) because updated MRI safety testing found they may heat up more than expected during MRI scans under certain conditions. The original instructions did not fully reflect these temperature increases.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck I.T.S. surgical screws for MRI safety
I.T.S. GmbH surgical screws (cannulated cancellous screws, cortical screws, and washers) used in orthopedic surgery can heat up more than expected during MRI scans under certain conditions. New safety testing found the temperature increases are higher than what was stated in the device instructions.
- 2026-04-01FDA-DeviceI.T.SClass IINationwideCheck MRI compatibility of I.T.S. surgical plates
I.T.S. GmbH orthopedic plates used in bone surgery were tested for safety during MRI scans. Testing found these plates can get hotter than originally stated when exposed to certain MRI conditions. The manufacturer has updated safety information to reflect the actual heating levels.