  • Boston Scientific Intracept instruments recalled for possible sterility breach

    The outer tray seal on certain Intracept access instruments may be breached, which may compromise sterility. The notice says a loss of sterility could lead to severe infection. No injuries have been reported.

    ModerateFDAJul 15, 2026Boston Scientific

  • ConvaTec Aquacel Ag Surgical SP dressing recalled for greater thickness

    ConvaTec, Inc. is recalling certain Aquacel Ag Surgical SP sterile surgical dressings (REF 413556, 9 cm x 15 cm) because they were manufactured out of specification, with greater thickness than intended. The recall covers 4,230 units in lot 5G00809, distributed nationwide. No injuries have been reported.

    ModerateFDAJul 15, 2026ConvaTec

  • Olympus single-use distal covers recalled for design replacement

    Olympus has released a newly designed version of its single-use Distal Cover (catalog MAJ-2315) to replace the previous design. The covers attach to EVIS EXERA III duodenovideoscopes (model TJF-Q190V) to cover the tip of the insertion tube and the area around the forceps elevator. The notice does not describe a specific hazard or report any injuries.

    ModerateFDAJul 15, 2026Olympus

  • Zimmer Biomet A.T.S 4000 tourniquet hoses recalled for sterility risk

    Zimmer Biomet is recalling two A.T.S 4000 TS tourniquet system hoses because the sealing area of the sterile pouch may be breached. If used, the hose may cause a moderate, localized infection. If found before use, surgery may be briefly extended while another device is located.

    ModerateFDAJul 8, 2026Zimmer Biomet2 products

  • GORE TAG thoracic endoprosthesis recalled for catheter separation

    W L Gore & Associates has recalled specific batches of the GORE TAG Thoracic Branch Endoprosthesis Side Branch Component due to a risk of catheter separation during use. This is a surgical implant used to repair certain aortic conditions. The separation could potentially affect the device's ability to function as intended.

    ModerateFDAJul 8, 2026Gore

  • Masimo Radius VSM devices recalled for false alarms, skin irritation

    Masimo is recalling two Radius VSM products: an ECG pre-connected electrode set and disposable blood pressure cuffs, covering 1,800 units in total. The ECG feature may incorrectly flag abnormally slow or fast heart rates in patients with normal rhythms, which could distract clinicians from true alarms and delay treatment. The blood pressure cuffs may have rough edges that can cause skin irritation or redness. No injuries have been reported.

    ModerateFDAJul 8, 2026Masimo Corporation2 products

  • Laborie Solar GI System recalled for moisture accumulation risk

    Laborie Medical Technologies' Solar GI System water perfusion devices may accumulate moisture inside tubing, filters, and pressure sensors during extended clinical use. This moisture buildup can affect how the system measures airflow and pressure, potentially impacting test results.

    ModerateFDAJul 8, 2026Laborie Medical Technologies Corp

  • Bard catheters and trays recalled for lidocaine ampoule concerns

    Bard Access Systems is recalling 14 products, including PICC, midline, and dialysis catheters and biopsy trays. The recall stems from lidocaine ampoules that were made under deficient manufacturing practices, and the supplier is recalling them. No injuries have been reported.

    ModerateFDAJul 8, 2026Bard14 products

  • BD Medline 111 gravity set recalled for engineering sample shipment

    BD Switzerland inadvertently shipped 50 engineering sample units of the Medline 111 Gravity Set with 3 Needle Free Y-Sites into the U.S. market. These units were labeled "Not for Human Use" and were not intended for patient care. No injuries have been reported.

    ModerateFDAJul 8, 2026BD

  • Covidien Shiley decannulation plug recalled for mislabeled DEHP

    The Shiley Disposable Decannulation Plug (DDCP), a red connector used on tracheostomy tubes to block airflow, carries labeling that incorrectly indicates it does not contain DEHP (a chemical compound). The product actually contains 30–35% DEHP. Users relying on the label may not be aware of DEHP exposure during use.

    ModerateFDAJul 8, 2026Covidien LLC

  • Zeiss Intrabeam radiation applicator recalled for manufacturing defect

    Certain Zeiss Intrabeam Spherical Applicator 1.5 cm units used in intraoperative radiation therapy did not meet manufacturing specifications. A total of 13 units were affected across the United States, Puerto Rico, and several other countries. No injuries have been reported.

    ModerateFDAJul 8, 2026Carl Zeiss Meditec

  • Siemens ARTIS icono ceiling monitor recalled for loose screws

    Siemens has identified a potential issue with the Display ceiling suspension (DCS Fix XL) component on ARTIS icono ceiling systems. Over time, screws that secure the monitor to the suspension may gradually loosen, which could affect the stability of the mounted display.

    ModerateFDAJul 8, 2026Siemens

  • RayCare 2024A SP1 recalled for beam note data loss risk

    RayCare 2024A SP1 oncology information system has a software issue where beam set delivery notes can lose data if modified at the same time or updated during an active treatment session. Users may not be notified when these changes occur, potentially affecting treatment records.

    ModerateFDAJul 8, 2026Raysearch Laboratories2 products

  • Medela ENFit ExSet extension set recalled for connector incompatibility

    Some Medela ENFit ExSet enteral extension sets contain incorrect connectors that may not attach properly to ENFit connections. This incompatibility could cause a slight delay in delivering nutrition or medication through the feeding tube.

    ModerateFDAJul 8, 2026Medela Inc

  • Accuray Cyberknife recalled for collimator fall risk

    In some cases, the collimator (a precision component) may not fully dock into the Cyberknife treatment system. This triggers a safety interlock that blocks motion from the operator console. However, if staff enter the treatment room and use the in-room control pendant to move the system, the collimator can release and fall from its housing, creating a potential impact hazard.

    ModerateFDAJul 8, 2026Accuray Incorporated

  • Siemens ARTIS icono systems recalled for loss of movement risk

    Siemens is recalling three models of its ARTIS icono imaging systems, 234 units in total, because the detector lift may stop moving after it is adjusted to its highest or lowest height limit. No injuries have been reported.

    ModerateFDAJul 8, 2026Siemens3 products

  • Accuray Cyberknife recalled for collimator drop risk

    In some cases, the collimator (a precision component) may not fully dock into the Cyberknife system. If operators bypass the safety interlock by using the in-room control pendant, the collimator can release and fall from the machine housing. This affects 48 Cyberknife units worldwide running software version 11X or earlier.

    ModerateFDAJul 8, 2026Accuray Incorporated

  • Pilling Wecksorb Neurosponges recalled for endotoxin contamination

    Pilling Wecksorb Neurosponges—sterile cotton sponges used during brain and spine surgery—may contain higher-than-safe levels of endotoxins (bacterial toxins). The affected sponges were manufactured between 2021 and 2024. No injuries have been reported, but elevated endotoxins could trigger an inflammatory response during surgery.

    ModerateFDAJul 8, 2026TELEFLEX LLC

  • Baxter operating table adapter recalled for loosening risk

    The Adapter for Head Positioning (Model 1739994) used with Baxter operating tables may become loose during patient positioning and move unintentionally. This affects 13 units distributed worldwide, primarily in Minnesota, Maryland, Pennsylvania, Canada, and numerous other countries.

    ModerateFDAJul 8, 2026Baxter

  • Boston Scientific CRE balloon catheters recalled for sterile packaging breach

    Boston Scientific is recalling 12 sizes of its CRE and CRE Pro Wireguided balloon catheters because the pouches they are packaged in may not keep the devices sterile. The recall covers 81,543 units per product and was distributed in the US and many other countries. No injuries have been reported.

    ModerateFDAJul 8, 2026Boston Scientific12 products

  • Hamilton Medical IntelliCuff recalled for motor malfunction risk

    The IntelliCuff portable cuff pressure controller may display a "Cuff System Leakage" error and cause the motor to pump continuously instead of maintaining the pressure set by the user. This can prevent the device from working as intended.

    ModerateFDAJul 8, 2026Hamilton Medical AG

  • Medtronic Harmony delivery catheter recalled for possible tip detachment

    Certain lots of the Medtronic Harmony Delivery Catheter System may have a manufacturing defect that increases the risk of the distal tip detaching from the delivery system during use. This could occur under specific procedural or anatomical conditions. About 1,881 units worldwide are affected.

    SeriousFDAJul 8, 2026Medtronic Heart Valves Division

  • Remel GC Agar Base recalled for reduced gonorrhea growth

    Remel, Inc. is recalling certain 100mm and 150mm GC Agar Base lab culture plates because some strains of gonorrhea bacteria may grow poorly or not at all on them. The recall covers specific lot numbers and was distributed in the US, Canada, Singapore, and the United Kingdom. No injuries have been reported.

    ModerateFDAJul 8, 2026Remel

  • Aligned Medical Solutions surgical kits recalled for bacterial contamination

    Windstone Medical Packaging is recalling four Aligned Medical Solutions surgical convenience kits, 1,525 units in total. The kits contain Webcol Large Alcohol Prep Pads that Cardinal Health recalled after a bacterium, Paenibacillus phoenicis, was found during a routine sterilization audit. The kits were distributed nationwide, with confirmed distribution in PA, GA, CA, and WA. No injuries have been reported.

    SeriousFDAJul 8, 2026Windstone Medical Packaging4 products

  • Medacta shoulder surgery broaches recalled for fracture risk

    Medacta's Short Monobloc Stem Broaches (sizes 7–16), used during shoulder replacement surgery, have a manufacturing issue that causes the surgical instruments to be positioned slightly off-center during the procedure. This misalignment may increase the risk of fractures in the upper arm bone.

    ModerateFDAJul 8, 2026Medacta Usa Inc

  • Depuy Synthes ATTUNE knee hinge components recalled for sterility risk

    DePuy (Ireland) is recalling 2 ATTUNE Revision Hinge knee components, a housing component and an insert component, because the outer sterile packaging was found stuck to the inner sterile packaging, which could compromise sterility. The recall covers 226 units distributed to Minnesota, North Carolina, and Texas. No injuries have been reported.

    ModerateFDAJul 8, 2026DEPUY (IRELAND)2 products

  • Medline reprocessed EP catheters recalled for particulate risk

    Specific lots of Medline reprocessed St. Jude Medical Response diagnostic electrophysiology catheters may contain small particles of residual material on surfaces that contact patients. If these particles dislodge during use, they could potentially cause blood clots, inflammation, or infection.

    SeriousFDAJul 8, 2026Medline

  • Mentor BD syringes recalled for expired product distribution

    A single lot of 10cc Luer Lok syringes (Lot 8298877) with an expiration date that had passed were distributed to hospitals and medical facilities in California, Florida, and Pennsylvania due to a failure in the distribution system's inventory controls. About 1,401 syringes were affected. No injuries have been reported.

    ModerateFDAJul 8, 2026Mentor

  • Medline kits with Swan-Ganz catheters recalled for leak risk

    Medline is recalling 8 convenience kits that contain Swan-Ganz catheters made by Becton Dickinson (BD). BD received complaints that the blue hub on these catheters can leak or break, which may lead to infection, medication loss, or blood loss. No injuries have been reported, and the kits were distributed in AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, and WI.

    ModerateFDAJul 8, 2026Medline8 products

  • Omnipod insulin pods recalled for insulin leak risk

    Insulet is recalling three Omnipod pod lines (Omnipod 5, Omnipod DASH, and Omnipod Eros), about 7 million units in total. Damage to the soft cannula during manufacturing can let insulin leak around the pod instead of being delivered, so users may get too little insulin. This can cause high blood sugar, and in severe cases prolonged high blood sugar can lead to diabetic ketoacidosis, hyperglycemic hyperosmolar syndrome, or death. No injuries have been reported.