Page 17 of 33- Hillrom VOLARA patient circuit kits recalled for nebulizer leakage
Four Hillrom VOLARA System patient circuit kits are included in this recall. Air and medication can leak from the nebulizer cup during therapy, which can lower a patient's oxygen level and make the treatment less effective. The problem has been linked to the nebulizer cup not being locked properly after medication is added during setup. No injuries have been reported.
SeriousFDAJul 8, 2026Baxter4 products
- Medline kits with Huons lidocaine recalled for quality issues
Medline is recalling 15 convenience kits that contain Huons lidocaine ampules. Huons issued the recall after an FDA inspection of its manufacturing site found quality problems with the ampules. No injuries have been reported.
ModerateFDAJul 8, 2026Medline15 products
- Hamilton Medical breathing circuit sets recalled for possible valve sticking
Certain coaxial breathing circuit sets used with Hamilton Medical ventilators (models C1, T1, and MR1) may have expiratory valve membranes that stick to the valve body. This defect is not always detectable before use and can trigger an "exhalation obstructed" alarm during ventilation, potentially leading to inadequate ventilation and oxygen desaturation in patients.
SeriousFDAJul 8, 2026Hamilton Medical AG
- Abiomed Impella CP introducer kits recalled for clot formation risk
Abiomed is recalling four Impella CP product sets that include 14 Fr Low Profile Introducer Kits, with 195 affected units in total. The introducer may form blood clots (thrombus) during prolonged use. No injuries have been reported, and the distribution covers the US states of FL, IL, MN, NJ, NY, and VA, plus Japan.
SeriousFDAJul 1, 2026Abiomed4 products
- Medline convenience kits recalled for unapproved design changes
Medline is recalling 16 convenience kits that contain 10mL polycarbonate colored syringes. The company found design changes to these kits that were not cleared by the FDA, and the products were not included in an earlier report. The kits were shipped nationwide and to Barbados, and no injuries have been reported.
ModerateFDAJul 1, 2026Medline16 products
- Boston Scientific INTERA 3000 pump recalled for drug leakage risk
Boston Scientific has recalled one INTERA 3000 Hepatic Artery Infusion Pump (model AP-0300H) due to the potential for leakage along the drug pathway from the pump through the catheter end. This is an implantable device used to deliver medication directly to the liver. The affected device was distributed in Ohio.
ModerateFDAJul 1, 2026Boston Scientific
- Philips Telemetry Monitor 5500 recalled for reset risk
Philips Telemetry Monitor 5500 (model 867232) devices may unexpectedly reset to a default state, causing them to lose their configuration settings and disconnect from the hospital's central monitoring system. This could prevent the monitor from reconnecting to the PIC iX central system until manually reconfigured.
ModerateFDAJul 1, 2026Philips
- Zimmer Persona Revision augment recalled for packaging error
Zimmer discovered that 12 units of the Persona Revision Trabecular Metal Femoral Distal Augment (Model 42-5572-066-10, Size 9/9+, 10mm thickness) contained an incorrect posterior augment component in the package instead of the correct distal augment. No patient injuries have been reported.
ModerateFDAJul 1, 2026Zimmer Biomet
- Medtronic Octopus tissue stabilizers recalled for tubing assembly error
Medtronic recalled three models of its Octopus tissue stabilizers, covering 1,012 units in total. During manufacturing, the company found that the tubing was attached backwards, with the short tube on the bottom of the canister and the long tube on top. No injuries have been reported.
ModerateFDAJul 1, 2026Medtronic3 products
- Dexcom G7 sensors recalled for infection and accuracy risks
Certain lots of Dexcom G7 continuous glucose monitoring sensors that were designated for destruction were stolen and later sold to pharmacies and distributors without authorization. Two affected lots have different risks: one was not properly sterilized and poses an infection risk, and the other may not provide accurate readings or detect dangerous blood sugar levels.
ModerateFDAJul 1, 2026Dexcom
- Surgify Halo surgical burrs recalled for breakage risk
Surgify Medical is recalling 19 Surgify Halo drill burr products used in bi-portal endoscopic spinal surgery because the burrs may break during use. The recall was filed as a Class II device recall, and no injuries have been reported. The products were distributed nationwide, including to MN, MA, IN, LA, CA, NY, and FL.
ModerateFDAJul 1, 2026Surgify Medical19 products
- IPG Medical Thuvera Laser Console recalled for false error display
A software issue in the Thuvera Laser Console (Software Version 2.2) causes the device to display a false error code 5018. This anomaly may lead clinicians to misinterpret the device status during operation.
ModerateFDAJul 1, 2026IPG Medical
- GE Healthcare AW Server 3.2 recalled for wrong patient images risk
GE Healthcare's AW Server 3.2 ext.6.5 has a software issue that can cause the system to display a previous patient's medical images instead of the correct patient's images when launching certain applications. There is no warning or error message when this happens. If a radiologist or clinician does not notice the error, they could misread images, make wrong diagnoses, or recommend incorrect treatment.
ModerateFDAJul 1, 2026GE
- Stryker INZONE coil detachment device recalled for battery failure
Stryker Neurovascular's INZONE Detachment System devices may experience premature battery drain, causing them to fail to power on, display faint indicators, or be unable to detach an embolization coil as intended. If a device fails during a procedure, medical staff may need to use an alternative technique to complete the intervention.
ModerateFDAJul 1, 2026Stryker
- Acumed Hexalob Screws recalled for breakage risk
Acumed Hexalob Screws used in wrist fixation systems may break during use due to a manufacturing error. The affected screws come in various sizes (2.7mm and 3.0mm) and have been distributed nationwide. No injuries have been reported.
ModerateFDAJul 1, 2026Acumed
- Zimmer Biomet Off-Axis shoulder reamer guides recalled for breakage
Zimmer Biomet is recalling five Off-Axis shoulder system reamer guide products distributed nationwide. The company received five complaints that the plastic guides caught during reaming and broke. No injuries have been reported.
ModerateFDAJul 1, 2026Zimmer Biomet5 products
- PFM Medical snare catheters recalled for tissue irritation risk
PFM Medical's Multi-Snare Set catheters (5 mm and 10 mm models) failed to meet biocompatibility safety standards. If used, the device material could cause localized inflammation and tissue irritation when in contact with blood vessels.
ModerateFDAJul 1, 2026PFM Medical
- Ventec VOCSN+Pro ventilator recalled for backup alarm failure
Ventec Life Systems' VOCSN+Pro respiratory ventilator may have a damaged backup battery that powers the alarm system. If primary power is lost and the backup battery is depleted or absent, the backup alarm may not sound to alert caregivers of the power loss.
SeriousFDAJul 1, 2026Ventec Life Systems
- TMJ Solutions bilateral implant recalled for component mismatch
TMJ Solutions Inc. has recalled one custom temporomandibular joint implant (model CHG020) that may contain incorrect patient-specific components. The mismatch caused a poor fit, resulting in a right-sided open bite for the affected patient.
ModerateFDAJul 1, 2026TMJ Solutions
- Philips Avalon fetal monitors recalled for speaker output loss risk
Philips is recalling two Avalon fetal monitor models, the FM20 and FM30, because the speaker connector may have been assembled incorrectly. Device cables can be pulled out of the housing, which may cause intermittent or permanent loss of speaker output. The recall includes 454 affected units, and no injuries have been reported.
ModerateFDAJul 1, 2026Philips2 products
- Intuitive Surgical Da Vinci recalled for E-brake failure risk
Certain Da Vinci surgical systems were distributed with a defective E-brake retainer pin. If the structural column fails, this pin can prevent the E-brake from engaging, potentially causing uncontrolled motion of the robotic arm. The pin may also loosen or back out, restricting arm movement or causing uncontrolled motion that could injure patient tissue.
SeriousFDAJul 1, 2026Intuitive Surgical
- Medline convenience kits recalled for anesthetic drug quality issues
Medline is recalling four groups of convenience kits, 7,738 units in total, because they include lidocaine and bupivacaine injections from supplier Arrow. Those injections were recalled after an FDA inspection of the supplier found quality issues. No injuries have been reported.
SeriousFDAJul 1, 2026Medline4 products
- HeartMate 3 system parts recalled for backup battery failures
Thoratec is recalling three HeartMate products: HeartMate 3 Left Ventricular System kits, System Controllers, and HM 11V Li-ion Backup Batteries. The 11-volt backup batteries can fail. No injuries have been reported.
SeriousFDAJun 24, 2026Thoratec3 products
- Merit Medical guide wires recalled for unsealed pouches
Merit Medical Systems' Splashwire Hydrophilic Guide Wire — used by doctors in vascular procedures — has been recalled because some product pouches were not sealed properly. Unsealed pouches can allow contamination or damage to the sterile device before use.
ModerateFDAJun 24, 2026Merit Medical
- Inspire respiratory and stimulation leads recalled for mislabeled Use By Date
Inspire Medical Systems is recalling two implant lead products, the Respiratory Sensing Lead and the Stimulation Lead, because they carry a Use By Date that goes beyond their validated three-year shelf life. The agency says material can degrade past the labeled date, which may lead to unplanned or premature revision surgery. No injuries have been reported.
ModerateFDAJun 24, 2026Inspire2 products
- BD Connecta stopcock recalled for loose connection risk
BD Connecta BD Luer-Lok 360 stopcocks used to administer IV fluids and medications may have a connection problem at Port C. When a plastic luer-lok syringe is attached, the connection may over-thread or continue spinning instead of tightening securely, making it feel unstable.
ModerateFDAJun 24, 2026BD
- Beta Bionics iLet Bionic Pancreas recalled for screen unlock failure
A firmware bug in the iLet Bionic Pancreas can cause the lock screen to fail to unlock when users slide the unlock slider. When this happens, users become stuck on the lock screen and cannot change insulin cartridges or enter meal information. The pump will continue delivering insulin until the cartridge empties and the battery drains, which may lead to high blood sugar.
ModerateFDAJun 24, 2026Beta Bionics
- Staar Surgical ICL Calculation Software recalled for diagram error
Staar Surgical's ICL Calculation Software Version 8.00 contains a coding error that can produce incorrect implantation orientation diagrams for toric intraocular lenses. The error occurs when multiple toric lens calculations are generated for the same eye between April 13–16, 2026. An incorrect diagram could lead to improper lens alignment during surgery.
ModerateFDAJun 24, 2026Staar Surgical
- Medline catheters and procedure kits recalled for performance defects
Medline recalled 30 products, including urethral catheters, catheterization trays, intermittent latex catheters, and medical procedure kits. During an internal review, the company found that some samples did not meet performance specifications, including surface irregularities, connectors that detached early from the catheter funnel, and some catheters with outer diameters outside specification. No injuries have been reported.
ModerateFDAJun 24, 2026Medline30 products
- Mito Red Light Super Mobile recalled for battery fire risk
The Mito Red Light Super Mobile near-infrared therapy device contains a lithium-ion battery that may malfunction and catch fire or burst into flames. The device has been distributed worldwide, including across the United States.
SeriousFDAJun 24, 2026Mito Red Light