Michigan recalls
3,531 federal recalls affecting Michigan in the last 12 months - 169 Michigan-specific plus 3,362 nationwide. Sourced from CPSC, FDA, USDA, and NHTSA.
- 2026-01-28FDA-DeviceEdan DiagnosticsClass IINationwideCheck your Edan fetal monitor model and serial number
The FDA identified potential cybersecurity issues with Edan Diagnostics fetal and maternal monitors (models F2 and F3). These bedside monitors are used during pregnancy and labor to track the health of pregnant women and their fetuses. The manufacturer has been notified and is working with the FDA to address the vulnerabilities.
- 2026-01-28FDA-DeviceCommunity ProductsClass IINationwideCheck your Rifton E-Pacer gait trainer for fraying
The body support strap on certain Rifton Equipment E-Pacer gait training devices (Model K660) may fray. Fraying could potentially affect the strap's function or safety during use. The company has identified specific serial numbers affected by this issue.
- 2026-01-28FDA-DeviceCommunity ProductsClass IINationwideCheck your Rifton TRAM lift device for strap damage
Rifton Equipment's Low Base TRAM powered lift device may have fraying on the body support strap used to secure clients during transfers. The fraying could potentially compromise the strap's ability to safely hold a person. This affects specific units manufactured between 2013 and 2015.
- 2026-01-28FDA-DeviceCommunity ProductsClass IINationwideCheck your Rifton TRAM lift device
Rifton Equipment's TRAM powered lift device (Model K310) may have fraying on the body support strap. This strap is used to support clients during seated transfers or sit-to-stand training. Fraying could potentially affect the strap's integrity during use.
- 2026-01-28FDA-DeviceSysmexClass IINationwideCheck your Sysmex TS-10/TS-10H Tube Sorter
Sysmex America's TS-10/TS-10H Tube Sorter may experience an unexpected loss of electromagnetic holding force in the lift mechanism under certain conditions. This could cause the lift to move downward unintentionally. The issue affects 44 specific units identified by serial number.
- 2026-01-28FDA-DeviceBeckman CoulterClass IINationwideStop using Beckman Coulter LDL Cholesterol OSR6x96
Beckman Coulter's LDL Cholesterol OSR6x96 reagent may give inaccurate test results when patients have high triglyceride levels (above 1500 mg/dL). The test can overestimate LDL cholesterol by up to 29% in some samples, which could affect patient care decisions.
- 2026-01-28FDA-DeviceIntuitive SurgicalClass IINationwideCheck your Ion Vision Probe Bag lot number
Intuitive Surgical has recalled certain Ion Vision Probe Bags (model 490127) because the sterile pouch seal may be compromised or incomplete. These bags are used to store the Vision Probe during lung biopsy procedures. A compromised seal could allow contamination of the sterile device.
- 2026-01-28FDA-DeviceMicroPort OrthopedicsClass IIStop using MicroPort HIPTURN hip implant trial
MicroPort Orthopedics distributed a hip implant trial component (HIPTURN FEM HEAD TRIAL 36MM) in the United States without FDA premarket authorization. The company sent approximately 91 units to distributors across multiple states. No injuries have been reported.
- 2026-01-28FDA-DeviceTyber MedicalClass IINationwideCheck Tyber Medical VOLT wrist plates
Tyber Medical's VOLT wrist treatment system left distal radius plates (model 02.426.691S, lot 291293) were manufactured with incorrect thread orientation. This defect could affect proper plate positioning and screw fixation during wrist surgery. Approximately 64 units were distributed nationwide.
- 2026-01-28FDA-DeviceTyber MedicalClass IINationwideCheck Tyber Medical VOLT wrist plates
Tyber Medical's VOLT Wrist Treatment System left wrist plates (Model 02.426.681S, Lot 291294) were manufactured with incorrect thread orientation. This defect could affect proper plate installation during wrist surgery. No injuries have been reported.
- 2026-01-28FDA-DeviceLeaseir TechnologiesClass IINationwideCheck your Leaseir MHR Xcell laser console label
Two Leaseir MHR Xcell surgical lasers used for hair removal and skin treatment were distributed with console labels missing the required "DANGER" symbol. The missing warning label could prevent users from recognizing laser hazard warnings. No injuries have been reported.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical angioplasty packs
Merit Medical Systems is recalling certain Allwell Angioplasty Packs because the inflation device handle may detach from the syringe during a procedure. This could interrupt treatment and require the procedure to be stopped or repeated. No injuries have been reported.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical Custom Procedure Kits
Merit Medical Systems is recalling certain Custom Procedure Kits because the inflation device handle may detach from the syringe during a medical procedure. This could interrupt the procedure or cause other complications. No injuries have been reported.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical Custom Manifold Kit
Merit Medical Systems is recalling its Custom Manifold Kit (model K09-13203A) because the inflation device handle may detach from the syringe during a medical procedure. This could interrupt treatment or require procedure restart. No injuries have been reported.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical inflation kits
Merit Medical Systems is recalling multiple Custom Inflation Kit models because the handle may detach from the syringe during use. This could interrupt a medical procedure. No injuries have been reported so far.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical vascular tray immediately
Merit Medical Systems is recalling specific lots of their Custom Waste Management Kit Vascular Tray because the inflation device handle may detach from the syringe during a medical procedure. This could interrupt patient care. No injuries have been reported.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using Merit Medical Allwell Inflation Device
Merit Medical Systems is recalling the Allwell Inflation Device (model IS-30-A) used in angiographic procedures because the handle may detach from the syringe during use. This could compromise the procedure and patient safety. The recall affects specific lot numbers distributed worldwide.
- 2026-01-28FDA-DeviceMerit MedicalClass IINationwideStop using recalled Merit Medical inflation device
Merit Medical Systems' basixCOMPAK inflation device handle may detach from the syringe during a medical procedure. This affects multiple reference numbers and lots of the 20 mL, 30 atm/bar model used in clinical settings.
- 2026-01-28FDA-DeviceMazor RoboticsClass IINationwideCheck Mazor X robotic system software
Mazor Robotics has recalled its Mazor X robotic guidance system due to software errors in certain versions that could cause incorrect positioning of surgical instruments during spinal surgery. The affected software versions are 5.0.1, 5.1.2, and 5.1.3. About 549 systems worldwide are affected, and no injuries have been reported.
- 2026-01-28FDA-DeviceBaxterClass IINationwideStop using recalled Baxter SIGMA Spectrum pump
Baxter Healthcare has recalled certain SIGMA Spectrum infusion pumps (model 35700BAX2) because they may contain defective grease on internal gears. This defective grease does not lubricate properly, which can cause premature wear of the pump mechanism and lead to the pump delivering too much medication or allowing medication to flow freely without stopping.
- 2026-01-28FDA-DeviceICU MedicalClass IINationwideContact ICU Medical about Plum Solo IV Pump
ICU Medical's Plum Solo and Duo Infusion pumps have a software issue that may prevent clinicians from programming an automatic line flush after piggyback therapy. In certain situations, the pump displays an error message that blocks the flush from being set up, even though the flush is a normal part of the treatment process.
- 2026-01-28FDA-DeviceICU MedicalClass IINationwideCheck ICU Medical Plum Duo IV Pump software
ICU Medical's Plum Duo Precision IV Pump may prevent clinicians from programming an automatic flush after piggyback therapy in certain situations. When this happens, the pump displays an error message and the flush cannot be set up, which could affect how medications are delivered.
- 2026-01-28FDA-DeviceICU MedicalClass IINationwideCheck your ICU Medical Plum Duo pump software
ICU Medical Plum Solo and Duo Infusion pumps have a software issue that may prevent clinicians from programming an automatic flush after piggyback therapy. When attempting to set up this flush feature, the pump may display an error message that blocks the programming, even though the flush itself is a safe feature.
- 2026-01-28FDA-DeviceGE HealthcareClass IINationwideCheck your GE Omni Legend PET scanner
Certain GE Healthcare Omni Legend PET scanners may occasionally produce streaking artifacts in scan images. This visual distortion appears as lines in the images displayed on the scanner console and could affect image quality used for clinical diagnosis.
- 2026-01-28FDA-DeviceICU MedicalClass IINationwideStop using ICU Medical Tego Connectors
ICU Medical's Tego Connector needle-free access devices have a silicone seal that may bulge, separate, or tear. This defect can cause fluid leaks, blockages, inability to draw or inject medication, and potential exposure to contamination or air in the bloodstream during hemodialysis or IV therapy.
- 2026-01-28FDA-DeviceRaysearch LaboratoriesClass IINationwideUpdate RayStation treatment planning software
RayStation versions 17.0.0 and 17.0.1 have a software error that prevents dose calculations from updating properly when certain regions of interest are modified. This can occur when regions have no contours but include material overrides or are marked as Bolus, Fixation, or Support types. Users may not see updated radiation doses after making changes.
- 2026-01-28FDA-DeviceRaysearch LaboratoriesClass IINationwideUpdate RayStation 2024B software immediately
RayStation version 2024B has a software issue where calculated radiation doses may not update correctly when certain regions of interest (ROIs) are modified. This affects ROIs without contours that have material overrides or are classified as Bolus, Fixation, or Support types. Clinics using this version should verify their dose calculations are accurate.
- 2026-01-28FDA-DeviceRaysearch LaboratoriesClass IINationwideCheck RayStation software versions 2024A SP1 and SP2
RayStation radiation therapy planning software versions 2024A SP1 and 2024A SP2 have a bug that prevents dose calculations from updating correctly when certain regions of interest (ROIs) are modified. This affects ROIs without contours that have material overrides or are classified as Bolus, Fixation, or Support types. If changes are made to these ROIs, the system may not recalculate doses as intended.
- 2026-01-28FDA-DeviceRaysearch LaboratoriesClass IINationwideUpdate RaySearch RayStation software immediately
RaySearch's RayStation radiation therapy planning software version 10B SP1 has a flaw where calculated radiation doses may not update correctly when certain regions of interest are modified. This affects regions with no contours that have material overrides or are classified as Bolus, Fixation, or Support types. Clinical staff may unknowingly rely on outdated dose calculations.
- 2026-01-28FDA-DeviceRaysearch LaboratoriesClass IINationwideCheck RayStation software for dose calculation issues
RayStation radiation therapy treatment planning software (versions 14.0.0, 15.0.0, and 15.1.3) may not properly update calculated radiation doses when certain regions of interest are modified. This can occur when regions have no contours but have material overrides or are classified as bolus, fixation, or support types. Changes to these regions may not trigger the dose recalculation as intended.