FDA medical device recalls
Formal FDA enforcement reports for medical devices — implants, monitors, diagnostic tests, and hospital and home-use equipment. 961 recalls in the last 12 months, in plain English.
Recent recalls
Page 18 of 33- GE Giraffe OmniBed infant incubators recalled for loose fasteners
GE HealthCare is recalling three Giraffe OmniBed product lines, about 43,300 units in total, used as infant incubators and warmers. Under certain conditions, such as impact during transport, fasteners holding the canopy soffit beneath the control panel can come loose, and a fastener could fall into the patient compartment. In rare cases, the canopy soffit could become loose or detach, which could potentially injure a patient. No injuries have been reported.
ModerateFDAJun 24, 2026GE3 products
- ConMed AirSeal surgical ports recalled for overpressure alert risk
ConMed is recalling 11 AirSeal access port and obturator products used in minimally invasive (laparoscopic) surgery. The recall is for a potential problem that could cause overpressure alerts during procedures. No injuries have been reported, and the products were distributed nationwide and internationally.
ModerateFDAJun 24, 2026ConMed11 products
- Howmedica Triathlon Baseplate Size 3 recalled for mislabeling
Stryker identified that some Triathlon Universal Baseplate Size 3 units (lot UZD9IB) were packaged in boxes labeled as Size 4, while Size 4 units (lot TYX7OB) were packaged in boxes labeled as Size 3. This labeling mix-up could result in using the wrong implant size during surgery.
ModerateFDAJun 24, 2026Howmedica Osteonics2 products
- Accriva directCHECK quality control recalled for incorrect labeling
Accriva Diagnostics' directCHECK ACT-LR whole blood quality control product (level 2, lot A6DLA001) has incorrect performance range information on its label. This labeling error could lead to misinterpretation of test results if the correct reference range is not known.
ModerateFDAJun 24, 2026Accriva Diagnostics
- Ritmed neurological sponges recalled for endotoxin contamination
AMD Medicom is recalling two Ritmed neurological sponge products used during brain and spine surgery. The sponges may not consistently meet the limit for endotoxin, a bacterial substance, in devices that contact cerebrospinal fluid. No injuries have been reported.
ModerateFDAJun 24, 2026AMD Medicom2 products
- Abiomed Impella CP Set recalled for low purge pressure risk
Abiomed's Impella CP Set with SmartAssist may experience low purge pressure issues that trigger repeated alarms. In some cases, this malfunction can interrupt or stop the device's mechanical heart support function.
SeriousFDAJun 24, 2026Abiomed
- Smith & Nephew TRIGEN humeral nail recalled for manufacturing defect
Smith & Nephew identified a manufacturing error in certain TRIGEN Proximal Straight Humeral Nails (8/7 X 16, Ref: 71760816) in which the distal holes are missing. If implanted without detection, the affected nail could require revision surgery. Five units were distributed internationally.
ModerateFDAJun 24, 2026Smith & Nephew
- Medtronic O-arm O2 recalled for image artifact risk
Medtronic has identified a firmware anomaly in the O-arm O2 Imaging System detector panel that can cause image artifacts during surgical imaging. The issue affects specific serial numbers of the O-arm O2 model BI70002000. No injuries have been reported.
ModerateFDAJun 24, 2026Medtronic
- Steris HarmonyAIR, ALYON surgical lights recalled over detaching buttons
Steris is recalling two surgical lighting systems, HarmonyAIR A-Series and ALYON, totaling about 2,650 units. The plus and minus buttons used to adjust light intensity on the light heads may detach. No injuries have been reported.
ModerateFDAJun 24, 2026Steris2 products
- Steris HarmonyAIR surgical lights recalled for paint chipping
Paint on the light heads of Steris HarmonyAIR A-Series Surgical Lighting Systems may chip or peel. This could potentially affect the lighting system's performance or introduce debris into the surgical field. The FDA has classified this as a Class II recall.
ModerateFDAJun 24, 2026Steris
- Medline convenience kits recalled for adhesive tubing defect
Medline is recalling 34 convenience and procedure kits because they contain a recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. The adhesive's butyrate tubing may crack during use, making the product unusable for application. No injuries have been reported.
ModerateFDAJun 24, 2026Medline34 products
- Philips Allura Xper and MultiDiagnost Eleva recalled for hand switch fault
Philips is recalling two imaging systems used in medical settings, the Allura Xper and MultiDiagnost Eleva, covering 2 product entries. The hand switch button may not fully release after use, which could cause loss of imaging function or unintended radiation exposure and extra contrast injection. No injuries have been reported.
ModerateFDAJun 24, 2026Philips2 products
- Arrow kits with lidocaine and bupivacaine recalled for sterility risk
Arrow is recalling 35 kits and sets that contain lidocaine, bupivacaine, or 0.9% sodium chloride from a supplier that was itself recalled for quality issues. If the sterility of these water-based injectable drugs is compromised, it could lead to serious or life-threatening injuries. No injuries have been reported, and the products were distributed nationwide.
SeriousFDAJun 24, 2026Arrow35 products
- Abiomed Impella Controller recalled for software error
A software error in certain Automated Impella Controller devices (used to manage left heart support blood pumps) may affect device function. The error affects controllers with software versions 6.0.1, 6.0.2, 6.0.3, 7.1 and above across multiple regions worldwide.
SeriousFDAJun 24, 2026Abiomed
- Giraffe and Panda infant warmers recalled for oxygen delivery risk
Datex-Ohmeda is recalling three infant resuscitation and warmer products, 550 units in total, because a blender knob shaft can loosen. This can cause the system to deliver an oxygen concentration that differs from the level set by the user, which could lead to too little or too much oxygen for the patient. Only units with a specific blender installed during service are affected.
SeriousFDAJun 24, 2026Datex-Ohmeda3 products
- Boston Scientific Orca valve recalled for suction button sticking
Boston Scientific is removing Orca Sterile Single Use Air/Water and Suction Valves from distribution because the suction button is sticking more often than expected. These valves are used during gastrointestinal endoscopic procedures to control air and water flow on endoscopes. No injuries have been reported.
ModerateFDAJun 17, 2026Boston Scientific
- Halyard medical kits recalled for sterilization concerns
AVID Medical is recalling four Halyard kit types because sealed pouches were distributed without validated sterilization. The kits may not be sterile, and the recall includes 476 kits total, distributed in Iowa. No injuries have been reported.
ModerateFDAJun 17, 2026AVID Medical4 products
- Elekta MOSAIQ cancer treatment software may fail to load plans
Elekta's MOSAIQ Oncology Information System software (version 3.2) may fail to load an adapted baseline shift plan during cancer treatment delivery if a communication error occurs. This could affect the accuracy of radiation therapy positioning. No injuries have been reported.
ModerateFDAJun 17, 2026Elekta
- Abbott Merlin PCS 3650 programmer software recalled for test termination malfunction
Abbott's Merlin PCS 3650 programmer software (versions 25.0.1 through 28.7.1) may malfunction during pacing capture threshold tests on pacemakers. When this occurs, the programmer may fail to successfully send the command to stop the test.
ModerateFDAJun 17, 2026Abbott
- Boston Scientific Clik X Anchor recalled for sterility concerns
Boston Scientific is recalling the Model SC-4318 Clik X Anchor (Lot #35107836) because sterile anchors may lack proper sterility assurance. Approximately 43 units were distributed across six U.S. states.
ModerateFDAJun 17, 2026Boston Scientific
- Boston Scientific CSK and TCN electrodes recalled for performance risk
Boston Scientific is recalling two lines of reusable electrodes, CSK and TCN, because they may not meet expected performance levels. The recall covers 2 products with about 81,000 units total, distributed across the US and several other countries. No injuries have been reported.
ModerateFDAJun 17, 2026Boston Scientific2 products
- Fresenius Bicarby Dialysate recalled for GTIN labeling error
Fresenius Medical Care's Bicarby Dialysate (Model RFP-400-G) has incorrect barcode labeling on some units. While the case label and printed information are correct, scanning the GTIN barcode alone could lead to identification errors. This could result in use of a dialysate with a potassium concentration different from what was prescribed.
ModerateFDAJun 17, 2026Fresenius Medical Care
- Advance Aluminum Chloride Gel recalled for wrong first aid insert
Some units of Inter-Med's Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit (catalog 504600-28) contain the package insert for Ferric Sulfate Gel instead. The wrong insert gives incorrect first aid instructions—specifically, it tells users to induce vomiting if the gel is swallowed, when the correct instructions say not to induce vomiting.
ModerateFDAJun 17, 2026Inter-Med
- DT MedTech Hintermann ankle implant recalled for wrong device packaged
DT MedTech discovered that some Hintermann Series Talar Implant units labeled as left implants actually contained right implants inside the packaging. This mix-up was found through customer complaints. A total of 67 units across five US states and two countries are affected.
ModerateFDAJun 17, 2026DT MedTech
- Siemens Dimension Tacrolimus Flex Reagent Cartridge recalled for result bias
Three lots of Siemens Dimension Tacrolimus Flex Reagent Cartridges show a potential positive bias in quality control and patient test results. This means the cartridges may report tacrolimus levels that are higher than actual values. The affected lots are FA6254, FA6260, and FA6286.
ModerateFDAJun 17, 2026Siemens
- The Binding Site EXENT Analyser recalled for QC bypass risk
The EXENT Analyser, an automated blood-test device, has a software issue where certain test plates were marked as passed without performing required quality checks. This could allow inaccurate M-protein measurements—either missed detections or wrong concentration values—to be reported to clinicians.
ModerateFDAJun 17, 2026The Binding Site
- Medartis APTUS compression screws recalled for misaligned hole
Medartis has recalled certain APTUS 5.0 headed cannulated compression screws (CCS) used in bone fixation surgery because the central hole in the screw is not properly centered. This affects 13 units across five states. No injuries have been reported.
ModerateFDAJun 17, 2026Medartis
- Sagemax NexxZr T dental discs recalled for fracture risk
Sagemax Bioceramics is recalling NexxZr T Multi A3.5 W98-16mm zirconium oxide discs used in dental crowns and bridges. Some discs may have a reduced or missing dentin layer, which can weaken them and cause fractures when patients chew or bite.
ModerateFDAJun 17, 2026Sagemax
- In2bones NeoSpan Compression Staple recalled for missing MRI label
In2bones USA has recalled NeoSpan SuperElastic Compression Staples used in hand and foot surgery because the required Magnetic Resonance Imaging (MRI) safety symbol is missing from the product packaging, inner trays, and patient labels. This labeling omission could prevent patients and healthcare providers from knowing whether the implant is safe to use during an MRI scan.
ModerateFDAJun 17, 2026In2bones
- Philips DigitalDiagnost C90 recalled for missing seismic kit
One Philips DigitalDiagnost C90 Flex/Value/Chest/ER X-ray system was shipped without the required Seismic kit for its Wall Stand VS2. The kit is designed to keep the stand stable and prevent damage during earthquakes or seismic events.
ModerateFDAJun 17, 2026Philips