Moderate riskFDA DeviceClass II

Medline convenience kits recalled for defective adhesive tubing that may crack

Medline convenience kits contain a recalled batch of Ferndale Laboratories MASTISOL Liquid Adhesive with defective tubing. The butyrate tubing in these kits may crack when used, making the adhesive unusable for medical application.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check your Medline kits for lot numbers 25JBU463 (ADD-A-CATH ARTNC185) or 25KBH082 (PICC DRSG TRAY DYNDC3291).
  2. Do not use affected kits. Contact your Medline representative or the FDA for replacement or disposal instructions.
  3. Report any problems with these kits to the FDA MedWatch program.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline convenience kits: ADD-A-CATH ARTERIAL INSERTION ARTNC185 PED SCHOOL AGE PICC DRSG TRAY DYNDC3291
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 34 products — see all of them.

Details

Units affected
49,654 kits total
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-2445-2026

In the FDA's words

Description
Medline convenience kits: ADD-A-CATH ARTERIAL INSERTION ARTNC185 PED SCHOOL AGE PICC DRSG TRAY DYNDC3291 Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application. ARTNC185 UDI-DI 10195327640729 Lot 25JBU463 DYNDC3291 UDI-DI 10195327334741 Lot 25KBH082 Distribution: US Nationwide distribution. OUS distribution pending. Affected quantity: 49,654 kits total
Hazard
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

34 products in this recall

Medline convenience kits recalled for adhesive tubing defect

Medline is recalling 34 convenience and procedure kits because they contain a recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. The adhesive's butyrate tubing may crack during use, making the product unusable for application. No injuries have been reported.

The agency lists each product separately. Check every one you might have.