- Philips X-ray ceiling carriages recalled for difficult vertical movement
Philips is recalling 6 imaging product lines whose ceiling suspension telescopic carriages were made between July 2024 and June 2025. The carriages may be hard to move up and down by hand, requiring extra force, because of a possible manufacturing issue. No injuries have been reported.
ModerateFDANov 19, 20256 products
- Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during video Urodynamics procedures. Component: N/A
Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.
FDANov 19, 2025Sonesta
- Abbott Eterna and Liberta RC implants recalled for communication loss
Abbott is recalling two implantable neurostimulator models, the Eterna SCS Model 32400 and the Liberta RC DBS Model 62400, because the devices may lose communication with the clinician programmer or patient controller. The recall covers 2 products, with about 21,437 units affected worldwide. No injuries have been reported.
ModerateFDANov 19, 20252 products
- GE Healthcare patient monitors recalled for battery loss risk
GE Healthcare is recalling five patient monitor products because certain batteries can lose the ability to hold a charge and provide backup power. If that happens while a monitor is running on battery, patient monitoring can stop, which could delay noticing a change in a patient's condition. No injuries have been reported.
ModerateFDANov 19, 20255 products
- Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
FDANov 19, 2025
- Medline anesthesia circuit kits recalled for tubing leak risk
Medline is recalling 10 listed kits that contain 120-inch expandable anesthesia circuit tubing, about 96,406 units in total. The tubing can crack, and the company received complaints of leaks, which may reduce oxygen and ventilation or interfere with anesthetic delivery, and may cause serious injury or death. Anesthetic gases may also escape into the operating room, and no injuries have been reported.
SeriousFDANov 19, 202510 products
- CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX100253, UNIFORM 3MM X 5.4 CM/FCX100305, UNIFORM 3.5MM X 6.6 CM/FCX100356, UNIFORM 4MM X 7.5 CM/FCX100407, UNIFORM 4MM X 11.5 CM/FCX100412, UNIFORM 5MM X 9.7 CM/FCX100509, UNIFORM 5MM X 17 CM/FCX100517, UNIFORM 6MM X 11.9 CM/FCX100611, UNIFORM 6MM X 26 CM/FCX100626, UNIFORM 7MM X 13.9 CM/FCX100713, UNIFORM 7MM X 30 CM/FCX100730, UNIFORM 8MM X 16.1 CM/FCX100816, UNIFORM 8MM X 29 CM/FCX100829, UNIFO
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
FDANov 19, 2025Cerenovus
- MAGNETOM Cima.X Upgrade. Model Number: 11689304.
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
FDANov 19, 2025
- Monarch Premium Kratom powders recalled for Salmonella risk
Lotus Vaping is recalling four Monarch Premium Kratom powders sold in mylar pouches from 0.5 oz to 8 oz: Bali Gold, Green Maeng Da, Red Bali and White Elephant. The products may be contaminated with Salmonella, and the recall covers products sold nationwide. No injuries have been reported.
SeriousFDANov 19, 20254 products
- Professional Botanicals capsules recalled for magnesium salicylate
Professional Botanicals is recalling three capsule products sold in 175cc plastic bottles: Nutra Calm (90 count), Total Comfort (60 count), and Tense-X (60 count). The FDA says the products contain magnesium salicylate, which is the reason for the recall. No injuries have been reported.
ModerateFDANov 19, 20253 products
- SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
FDANov 12, 2025
- B-K Medical ultrasound systems recalled for battery instruction gaps
B-K Medical is recalling six ultrasound system models, with about 7,800 units affected worldwide. The user guide and service manual do not clearly state battery replacement intervals, end-of-life handling, or general battery safety precautions. No injuries have been reported.
ModerateFDANov 12, 20256 products
- Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell¿ Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
FDANov 12, 2025
- Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm O
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
FDANov 12, 2025Total Joint Orthopedics
- Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
FDANov 12, 20252 products
- Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi,
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
FDANov 12, 2025
- MAHURKAR triple lumen catheter trays recalled for seal integrity risk
Mozarc Medical is recalling six MAHURKAR Acute High Pressure Triple Lumen Catheter Tray products. The header bag packaging may have a seal integrity problem. No injuries have been reported.
ModerateFDANov 12, 2025Mozarc Medical6 products
- TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
FDANov 12, 2025
- Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
FDANov 12, 20252 products
- Brand Name: Inclusive Product Name: Inclusive Titanium Screw compatible with: Biomet 3i" Certain 3.4/4.1/5.0/6.0 mm Model/Catalog Number: 70-1047-COM0117
Incorrect titanium screw, packaged with dental implant.
FDANov 12, 2025Prismatik Dentalcraft
- Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
FDANov 12, 2025Spacelabs Healthcare
- CODMAN Disposable Perforator 11mm. Cranial Perforator.
Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.
FDANov 12, 20252 products
- Acetaminophen 500mg Caffeine 65mg caplets, packaged as 2 caplets per packet further packaged in a 50-count box, AERO TAB, Manufactured for AERO HEALTHCARE US, Valley-Cottage, NY, 10989, NDC 55305-135-01
Labeling: Label Mix-up. This issue affects the outer box labeling only. The box incorrectly states the ingredients Acetaminophen 500mg and Caffeine 65mg. The inner pouch correctly states the ingredients are Aspirin (NSAID)*500mg and Caffeine 32.5mg.
FDANov 12, 2025Aero Healthcare
- clomiPRAMINE Hydrochloride, Capsules, USP, 25 mg, packaged in a) 30-count bottles (NDC 16714-849-01), b) 90-count bottles (NDC 16714-849-02), c) 100-count bottles (NDC 16714-849-03), Rx only, Manufactured for: Northstar Rx LLC, Memphis, TN 38141, Manufactured by: Zydus Lifesciences Ltd., Ahmedabad, India
cGMP deviations: an observed Out of Specification of Nitrosamine Drug Substance-Related Impurities (NDSRIs), N-Nitroso Desmethyl-Clomipramine, above the FDA acceptable intake limit.
FDANov 12, 20253 products
- Amoxicillin and Clavulanate Potassium for Oral Suspension USP, 200 mg/28.5 mg per 5mL, 100 mL (when reconstituted), Rx only, manufactured in Canada by: Teva Canada Limited, Toronto, Canada M1B 2K9; Manufactured For: Teva Pharmaceuticals USA, Inc., Parsippany, NJ 07054 NDC 0093-2277-7.
Subpotent drug; Clavulanate Potassium component
FDANov 12, 2025
- Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
FDANov 5, 2025Microgenics
- Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique
As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.
FDANov 5, 2025
- Good Neighbor Pharmacy, TRUE METRIX Self Monitoring Blood Glucose Meter. Model/Catalog Number: RE4011-01.
Manufacturing defect causing LCD display issues with missing or partial segments (ghosting) that can result in misreading glucose values, potentially leading to inappropriate treatment decisions for diabetes management.
FDANov 5, 2025Trividia Health
- CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
FDANov 5, 2025
- MST ArcDUO 9mm, REF: ARC-0002, single-use, sterile ophthalmic knife, used with Suction Set, Part: 48-1000-01
An ophthalmic knife has a suction set (not direct patient-contacting), which contains springs contaminated with an unidentified residue that may potentially cause eye irritation
FDANov 5, 2025MicroSurgical Technology