- Check your oxycodone tablets for loose packaging
American Health Packaging is recalling certain oxycodone hydrochloride 5 mg tablets because the blister card seals are defective or missing. This allows tablets to fall out of their compartments. The recall affects 31,676 packages distributed nationwide with specific lot numbers and expiration dates.
ModerateFDAFeb 18, 2026
- Do not use Fresenius Kabi ketamine HCl injection
Fresenius Kabi Compounding has recalled five specific lots of ketamine HCl injection (10 mg per mL) distributed nationwide because the manufacturer cannot confirm the product remains sterile. The affected lots were produced between February and April 2026.
ModerateFDAFeb 18, 20265 products
- Stop using Asteria Health estradiol pellets
Asteria Health is recalling estradiol hormone pellets (10 mg) because they may contain metal particles. The affected pellets were distributed nationwide. No injuries have been reported so far.
ModerateFDAFeb 18, 2026Asteria Health25 products
- Stop using Unichem Doxazosin 4 mg tablets
Unichem Pharmaceuticals distributed 60,000 tablets of Doxazosin 4 mg with incorrect imprinting on the tablets themselves. This labeling error could cause confusion about the medication's identity and strength. The affected lot is GDSH25006, which expires in August 2027.
ModerateFDAFeb 18, 2026Unichem Pharmaceuticals
- Pro Numb tattoo numbing sprays recalled for manufacturing quality issues
Pro Numb is recalling two tattoo numbing sprays because they were made in ways that did not meet FDA manufacturing standards. The recall covers Pro Numb Tattoo Numbing Spray 5% Lidocaine and Pro Numb Tattoo Numbing Spray For Sensitive Skin 4% Lidocaine, sold nationwide in the USA. No injuries have been reported.
ModerateFDAFeb 18, 2026Pro Numb2 products
- Molift mobile lifts and sling bars recalled for bolt failure risk
Etac is recalling four Molift mobile patient hoist and sling bar products, covering 255 affected units. A customer reported that the bolt mounting the lifting bar broke during assembly and load testing. The company found that material deformation between the lifting bar and the device arm can limit joint mobility and stop the equipment from working properly. No injuries have been reported.
ModerateFDAFeb 11, 2026Etac4 products
- Check Maquet bubble sensor cable connections
Maquet Cardiopulmonary identified a durability problem with the connecting cable on its Venous Bubble Sensor (Model 701055720). Excessive bending of the cable can damage the sensor or loosen the connection, causing error messages on the connected medical device—either temporarily when the cable moves or permanently if the connection fails.
ModerateFDAFeb 11, 2026
- Stop using affected FUJIFILM FDR Visionary Suite units
FUJIFILM Healthcare Americas has recalled certain FDR Visionary Suite X-ray systems (Model CH-200) because internal electrical contacts can become welded together. This may allow power to continue flowing to the system even after it is turned off, which could cause the circuit to overheat, damage nearby parts, and create smoke.
ModerateFDAFeb 11, 2026
- Check Wilson-Cook nasal feeding tubes for missing part
Wilson-Cook Medical is recalling certain nasal jejunal feeding tubes (model NJFT-10) because some units were shipped without the nasal transfer tube component, making them incomplete and unusable. The recall affects 60 units distributed in Australia.
ModerateFDAFeb 11, 2026
- CleverCut and Olympus sphincterotomes recalled for deformation risk
Olympus is recalling 30 sphincterotome devices, which are used with an endoscope during medical procedures. Some units were not thermoformed as required and could deform, losing performance. No injuries have been reported.
ModerateFDAFeb 11, 202630 products
- LotFancy, Vivoo and URIMED urine test strips recalled over unauthorized marketing
Changchun Wancheng is recalling 11 product listings of home test strips and diagnostic tests, including urinary tract infection, ovulation, menopause, fertility and Vivoo tests. The devices were distributed in the U.S. before FDA clearance (510(k)), and they may give false or inaccurate results to users, which could lead to inappropriate medical treatment. The recall was distributed nationwide, with confirmed distribution in Texas, Georgia and California.
ModerateFDAFeb 11, 2026Changchun Wancheng11 products
- Vortex Surgical eye surgery tools recalled for sterility risk
Vortex Surgical is recalling 13 ophthalmic surgical products, including laser probes, forceps, cannulas, and related kits, because the Tyvek pouch seals may have voids. A damaged seal can compromise the sterile barrier and allow bioburden contamination, which could lead to infection. No injuries have been reported.
ModerateFDAFeb 11, 2026Vortex Surgical13 products
- Stop using Focalyx Fusion software
Focalyx Fusion, a software device used to guide prostate biopsies and diagnostic procedures, may not work properly on Windows 10 operating systems. This malfunction could harm patients during biopsy or delay needed treatment. The manufacturer recommends stopping use of affected software versions until the device is confirmed to work with Windows 11.
ModerateFDAFeb 11, 2026Focalyx
- Siemens CT scanner software recalled for lacking FDA clearance
Siemens is recalling three brain imaging software applications installed on certain CT systems because they have not received FDA 510(k) clearance. The recall covers 4 CT system models, with 97 units affected, distributed in the US and Canada, Costa Rica, India, and Israel. No injuries have been reported.
ModerateFDAFeb 11, 20264 products
- Check Tornier orthopedic screws for size mismatch
A specific lot of Tornier surgical screws (catalog DWJ318, lot AZ4824290) is labeled as 5.0mm×18mm but actually contains 5.0mm×14mm screws. If the wrong size is used without detection during surgery, it could increase the risk of mechanical failure and the need for revision surgery.
ModerateFDAFeb 11, 2026Tornier
- Stop using affected Beckman Coulter CellMek systems
A software error in the Beckman Coulter CellMek SPS Sample Preparation System (model C44603, software version 2.3.90.0) causes samples to be dispensed incorrectly, potentially damaging the instrument and losing 60–100 microliters of sample fluid. This could lead to inaccurate test results and incorrect patient care decisions.
ModerateFDAFeb 11, 2026
- Encore Medical EMPOWR implants recalled for packaging discrepancy
Encore Medical is recalling two EMPOWR surgical device products, 68 units in total, because some packages may contain the wrong acetabular system or may have a packaging discrepancy. One acetabular system package was found to contain a different device, and other acetabular system and knee insert devices may be affected. The problem could delay surgery while devices are exchanged or lead to a revision surgery. No injuries have been reported.
ModerateFDAFeb 11, 2026Encore Medical2 products
- Check da Vinci surgical system instrument arms
A software error allowed robotic instrument arms with bent or misaligned parts to remain in use during surgery. These parts can wear out and potentially fail during a procedure. Intuitive Surgical identified specific serial numbers of affected systems.
ModerateFDAFeb 11, 2026
- Prismaflex and Oxiris dialysis sets recalled for detachment risk
Vantive is recalling eight Prismaflex and Oxiris dialysis set products distributed nationwide. The deaeration chamber on these sets may become dislodged from the Prismaflex Control Unit. No injuries have been reported.
ModerateFDAFeb 11, 2026Vantive8 products
- Stop using Olympus KD-640L electrosurgical knives
Olympus Corporation is recalling certain single-use electrosurgical knives (model KD-640L) used during endoscopic surgery. The cutting knife can overheat, burn, and break during use, which could affect the surgery. No injuries have been reported.
ModerateFDAFeb 11, 20262 products
- Stop using recalled Beckman Coulter diluent lots
Beckman Coulter has recalled specific lots of COULTER DxH Diluent and COULTER DxH ECO Diluent used in blood cell analysis systems. These affected lots can produce inaccurate platelet count readings during daily quality checks, which may lead to unreliable test results.
ModerateFDAFeb 11, 2026
- Stop using Sun Pharmaceutical Diclofenac Sodium Gel
Sun Pharmaceutical is recalling one lot of Diclofenac Sodium Topical Gel, 3% (Lot #AD92721, expiring 3/31/2027) because the gel's thickness is slightly lower than required specifications. The product was distributed nationwide and is sold by prescription only.
ModerateFDAFeb 11, 2026
- Stop using Granules trazodone 50 mg
Granules Pharmaceuticals is recalling certain bottles of trazodone 50 mg because some bottles may contain foreign tablets or capsules mixed in. This affects 71,424 bottles with batch number 6160008A and an expiration date of 12/31/2026. No injuries have been reported.
ModerateFDAFeb 11, 2026Granules
- Stop using Unichem bisoprolol/hydrochlorothiazide tablets
Unichem Pharmaceuticals is recalling one lot of bisoprolol fumarate and hydrochlorothiazide tablets (2.5 mg/6.25 mg) because they contain higher-than-allowed levels of N-Nitroso Bisoprolol, a potentially harmful impurity. The affected lot was manufactured in India and distributed nationwide.
ModerateFDAFeb 11, 2026Unichem Pharmaceuticals
- Check BEAR Implant Model 1000 expiration date
Miach Orthopaedics BEAR Implant Model 1000 units have an incorrect expiration date printed on the label that extends beyond the product's actual shelf life. Because the expiration date is wrong, the UDI (Unique Device Identifier) is also incorrect. This affects 284 units distributed nationwide.
ModerateFDAFeb 4, 2026Miach Orthopaedics
- Stop using LSL Healthcare Central Line Dressing Kit
LSL Healthcare's Central Line Dressing Kit (Model 2519CP) contains BD ChloraPrep Triple Swabsticks with damaged packaging—some applicator seals are open. An open seal can allow the sterile applicator to become contaminated before use. This affects about 1,400 kits distributed in Idaho with lot number 5G3119 and expiration date 7/31/26.
ModerateFDAFeb 4, 2026LSL Healthcare
- Check LimFlow Vector device expiration date
LimFlow Vector vascular devices (Model VT-US-23, Lot 82321838) were distributed with an incorrect expiration date printed on the packaging. This affects 33 units distributed across multiple U.S. states. An incorrect expiration date could lead to use of a device beyond its safe lifespan.
ModerateFDAFeb 4, 2026LimFlow
- Check your BD Alaris Pump Module 8100
Some BD Alaris Pump Module 8100 infusion pumps have a software date/time issue related to daylight saving time that may prevent them from connecting to hospital networks. This affects only the network connectivity function; the pump's ability to deliver medication is not impacted. The issue involves 189 devices distributed in Arizona, Indiana, Ohio, and Virginia.
ModerateFDAFeb 4, 2026
- Philips Ingenia and Evolution MRI systems recalled for stiffness errors
Philips is recalling 13 MR imaging system products because exported MR Elastography stiffness maps may show incorrect stiffness values when viewed in a PACS image viewer. No injuries have been reported.
ModerateFDAFeb 4, 202613 products
- Check your Zimmer tourniquet system software
The screen on Zimmer A.T.S. 5000TS Tourniquet Systems may freeze or stop responding when used in certain non-English language settings. This can delay surgery while staff locate an alternate device. The issue affects systems running software version v2.04 or earlier.
ModerateFDAFeb 4, 20262 products